FAQ - Frequently Asked Questions
- Products Search > HPLC > System Controller
Keyword Search
Frequently Referenced FAQ
FAQ within [ System Controller ]
-
«Main Cause» The column oven did not stabilize at the set temperature during validation and calibration. «Countermeasure» Check whether there are any probl... Show Detail
- No:3534
- Date of Issue:2020/12/02 16:11
-
«Main Cause» A rack number that is different from the rack number set in the sequence may have been notified to the instrument. «Countermeasure» Clear the e... Show Detail
- No:3533
- Date of Issue:2020/12/02 16:11
-
«Main Cause» A vial number that is different from the sample vial number set in the sequence may have been notified to the instrument. «Countermeasure» Clea... Show Detail
- No:3532
- Date of Issue:2020/12/02 16:11
-
«Main Cause» The injection volume setting value exceeds the [Max Injection Volume] setting in the system settings. «Countermeasure» Change the injection vol... Show Detail
- No:3531
- Date of Issue:2020/12/02 16:11
-
«Main Cause» More than 10 instruments are connected to the same group. «Countermeasure» Check the number of instruments connected to the instrument group wh... Show Detail
- No:3530
- Date of Issue:2020/12/02 16:11
-
«Main Cause» Purge was executed because the replacement pump pressure reached the [minimum pressure] in the method parameters. «Countermeasure» Inspect for ... Show Detail
- No:3529
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The system check is timed out. «Countermeasure» Execute system check again. If the error is not resolved, contact a Shimadzu sales or service r... Show Detail
- No:3528
- Date of Issue:2020/12/02 16:10
-
«Main Cause» IP address renewal/refresh failed. «Countermeasure» Contact the network administrator. Set the IP address settings (automatic acquisition/manua... Show Detail
- No:3527
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The IP address is set to "192.168.200.96" because the instrument failed to automatically obtain an IP address. «Countermeasure» Contact the ne... Show Detail
- No:3526
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The buffer memory used for transmitting data from the instrument to the workstation is full. «Countermeasure» Check the network connection stat... Show Detail
- No:3525
- Date of Issue:2020/12/02 16:10
111-120 of Total 150