• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ System Controller ]

101-110 of Total 150

≪ 11 / 15page ≫
  • Error Code:[0x1001]

    «Main Cause» While full-regenerating, activate on or switch cartridge command was operated. «Countermeasure» Do not operate those commands while full regene... Show Detail

    • No:3549
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x9063]

    «Main Cause» An error occurred in communication between the autosampler or column oven and controller, but the error recovered. «Countermeasure» Clear the ... Show Detail

    • No:3543
    • Date of Issue:2020/12/02 16:11
  • Error Code:[0x9062]

    «Main Cause» An error occurred in communication between the autosampler or column oven and controller, and the communication was disconnected. «Countermeas... Show Detail

    • No:3542
    • Date of Issue:2020/12/02 16:11
  • Error Code:[0x9061]

    «Main Cause» An error occurred in communication between the replacement pump and controller, but the error recovered. «Countermeasure» Clear the error. Show Detail

    • No:3541
    • Date of Issue:2020/12/02 16:11
  • Error Code:[0x9060]

    «Main Cause» An error occurred in communication between the replacement pump and controller, and the communication was disconnected. «Countermeasure» Clear ... Show Detail

    • No:3540
    • Date of Issue:2020/12/02 16:11
  • Error Code:[0x907E]

    «Main Cause» When overlapping injection is run, the sample vial is not set in the sample vial number specified for the next acquisition. «Countermeasure» C... Show Detail

    • No:3539
    • Date of Issue:2020/12/02 16:11
  • Error Code:[0x907D]

    «Main Cause» An error occurred in communication between the detector and controller. «Countermeasure» Clear the error. If the error occurs again, contact a... Show Detail

    • No:3538
    • Date of Issue:2020/12/02 16:11
  • Error Code:[0x907C]

    «Main Cause» The instrument power was turned OFF during acquisition. «Countermeasure» Check whether the power was turned OFF via the instrument console or t... Show Detail

    • No:3537
    • Date of Issue:2020/12/02 16:11
  • Error Code:[0x907B]

    «Main Cause» The sample cooler did not stabilize at the set temperature during validation and calibration. «Countermeasure» Check whether there are any pro... Show Detail

    • No:3536
    • Date of Issue:2020/12/02 16:11
  • Error Code:[0x907A]

    «Main Cause» There is no sample vial set at the sample vial number specified in the sequence. «Countermeasure» Set a sample vial at the sample vial number s... Show Detail

    • No:3535
    • Date of Issue:2020/12/02 16:11

101-110 of Total 150

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.