Frequently Referenced FAQ

FAQ within [ System Controller ]

101-110 of Total 150

≪ 11 / 15page ≫
  • Error Code:[0x4001]

    «Main Cause» Because of communication error, communication with the workstation could not be established. «Countermeasure» Check the communication environme... Show Detail

    • No:3801
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x4004]

    «Main Cause» Because of communication error, communication with the workstation could not be established. «Countermeasure» Check the communication environme... Show Detail

    • No:3802
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x8001]

    «Main Cause» There was no response from a instrument for a given time during a system check. «Countermeasure» Check the instrument connection status and per... Show Detail

    • No:3803
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x8002]

    «Main Cause» The date and time specified in the startup has passed the present date and time. «Countermeasure» Specify a future date/time as the startup dat... Show Detail

    • No:3804
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x8003]

    «Main Cause» Solvent delivery was stopped because the pressure exceeded the system upper pressure limit value. «Countermeasure» Change the syste... Show Detail

    • No:3805
    • Date of Issue:2020/12/02 16:43
    • Last Updated:2020/12/04 18:40
  • Error Code:[0x8010]

    «Main Cause» The pump was purged because pressure did not reach the lower pressure limit value, but an error occurred again, so solvent delivery was stopped.... Show Detail

    • No:3806
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x8084]

    «Main Cause» A separate workstation than connected to the system controller is connected to the PDA detector. (System controllers and PDA detectors with an e... Show Detail

    • No:3807
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80E0]

    «Main Cause» Eleven or more system controllers and PDA detectors are assigned to the same group. «Countermeasure» Specify ten or fewer devices per group. Show Detail

    • No:3808
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80F0]

    «Main Cause» Multiple PDA detectors attempted to connect to the same system controller (redundant parent-child relationships). «Countermeasure» Change PDA d... Show Detail

    • No:3809
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80F3]

    «Main Cause» The IP address for the PDA detector was changed during use. «Countermeasure» The parent-child relationship might have failed between the system... Show Detail

    • No:3810
    • Date of Issue:2020/12/02 16:43