• Font Size
  • S
  • M
  • L

FAQ within [ Supercritical Fluid Chromatograph ]

11-20 of Total 34

2 / 4page
  • Error Code:[0x8B41] SFE ERR T.MAX H*

    «Main Cause» The temperature of the extraction vessel exceeded the [Maximum Temperature] of the instrument parameters. «Countermeasure» Maximum temperature... Show Detail

    • No:3695
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B42] SFE ERR TEMP SENS H*

    «Main Cause» The extraction vessel temperature sensor detection value is abnormal. The CO2 gas sensor may be disconnected. «Countermeasure» Press the (powe... Show Detail

    • No:3696
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B43] SFE ERR T.MAX TUBE

    «Main Cause» The temperature of the pre-heater exceeded the [Maximum Temperature] of the instrument parameters. «Countermeasure» Maximum temperature must b... Show Detail

    • No:3697
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B44] SFE ERR TEMP SENS TB

    «Main Cause» The pre-heater temperature sensor detection value is abnormal. The pre-heater temperature sensor may be disconnected. «Countermeasure» Press th... Show Detail

    • No:3698
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B45] SFE LOW SET TEMP H*

    «Main Cause» Set temperature value is inappropriate. Temperature is specified with lower than the room temperature plus 5ºC. «Countermeasure» Set temperatur... Show Detail

    • No:3699
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B46] SFE ERR LEAK VESSEL*

    «Main Cause» The temperature of the extraction container has decreased. Check if the extraction pack is tightened. «Countermeasure» Please confirm the extra... Show Detail

    • No:3700
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B47] SFE ERR HEATER H*

    «Main Cause» The extraction pack cell temperature control heater has failed. «Countermeasure» Press the (power button) on the instrument to turn the power O... Show Detail

    • No:3701
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B48] SFE ERR HEATER TUBE

    «Main Cause» The preheater is abnormal. «Countermeasure» Press the (power button) on the instrument to turn the power OFF and contact a Shimadzu sales or se... Show Detail

    • No:3702
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B49] SFE ERR TEMP SENS VE

    «Main Cause» The vessel thermosensor is abnormal. «Countermeasure» Press the (power button) on the instrument to turn the power OFF and contact a Shimadzu s... Show Detail

    • No:3703
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B50] SFE ERR TEMP SENS RM

    «Main Cause» Room temperature sensor is abnormal. «Countermeasure» Press the (power button) on the instrument to turn the power OFF and contact a Shimadzu s... Show Detail

    • No:3704
    • Date of Issue:2020/12/02 16:31

11-20 of Total 34

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.