(LC) Cleaning the extraction vessel (SFE)
Cleaning step after use (reference) 1. Disassemble the extraction vessele and take out the contents. HINT: If the contents do not come out... Show Detail
- No:6926
- Date of Issue:2021/09/08 10:22
-
(LC) Extraction vessel open/close kit (Option) (SFE)
1. Mount the fixing board of theextraction vessel opening and closingkit on the desk. (1) Fixing board of extraction vesselopening and closing kit, ... Show Detail
- No:6921
- Date of Issue:2021/09/08 09:34
- Last Updated:2021/09/08 10:23
- Brached FAQ
-
•Diaphragm, BPR Block(Click) •Needle, Needle Seal(Click) •Stator, Rotor(Click) •Suction Filter(Click) •Extraction Vessel Filter, ... Show Detail
- No:6558
- Date of Issue:2021/07/12 08:53
- Last Updated:2024/09/30 13:59
-
Error Code:[0x8BB1] SFC ERR LEAK SENS
«Main Cause» The leak sensor detection value is abnormal. The leak sensor may be disconnected. «Countermeasure» Press the (power button) on the instrument ... Show Detail
- No:3715
- Date of Issue:2020/12/02 16:31
-
Error Code:[0x8BB0] SFC ERR LEAK DETECT
«Main Cause» Eluent leak can be detected in the BPR. «Countermeasure» Stop the pump flow and then execute any action to stop the eluent leak. See detail In... Show Detail
- No:3714
- Date of Issue:2020/12/02 16:31
-
Error Code:[0x8BA2] PUMP HEAD COOLER ERR COOL PUMP
«Main Cause» The coolant pump for cooling the pump head has failed. «Countermeasure» Press the (power button) on the instrument to turn the power OFF and c... Show Detail
- No:3713
- Date of Issue:2020/12/02 16:31
-
Error Code:[0x8BA1] PUMP HEAD COOLER ERR COOLER FAN
«Main Cause» The pump head cooling fan is not rotating correctly. «Countermeasure» If anything is interfering with the fan rotation, turn the power OFF and... Show Detail
- No:3712
- Date of Issue:2020/12/02 16:31
-
Error Code:[0x8BA0] PUMP HEAD COOLER ERR TEMP SENS
«Main Cause» Pump head temperature sensor is open or short-circuit. «Countermeasure» Press the (power button) on the instrument to turn the power OFF and c... Show Detail
- No:3711
- Date of Issue:2020/12/02 16:31
-
Error Code:[0x8B95] SFC ERR TEMP SENS
«Main Cause» Lamp housing temperature sensor is open or short-circuit. «Countermeasure» Press the (power button) on the instrument to turn the power OFF an... Show Detail
- No:3710
- Date of Issue:2020/12/02 16:31
-
Error Code:[0x8B94] SFC BPR COVER OPEN
«Main Cause» BPR cover is open. «Countermeasure» Attach the BPR cover securely. Show Detail
- No:3709
- Date of Issue:2020/12/02 16:31