FAQ - Frequently Asked Questions
- Products Search > HPLC > Solvent Delivery Module
Keyword Search
Frequently Referenced FAQ
FAQ within [ Solvent Delivery Module ]
-
Error Code:[0x9116] ERROR OPEN DRAIN VALVE
«Main Cause» Solvent delivery was stopped because pressure became high during purging. «Countermeasure» If an error message is displayed during auto-purging... Show Detail
- No:3877
- Date of Issue:2020/12/02 16:44
-
Error Code:[0x9117] FATAL INSIDE FAN
«Main Cause» A motor cooling fan rotation problem was detected. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service rep... Show Detail
- No:3878
- Date of Issue:2020/12/02 16:44
-
The rinse solution volume in the seal automatic rinsing k...
Is there leakage from the plunger seal? Replace the plunger seal. If the plunger seal service life was extremely short, then also replace the plunger. Show Detail
- No:4249
- Date of Issue:2020/12/03 09:24
-
Solvent Flow Does Not Stabilize
Manually purge the flow channels, degas the mobile phase, clean the suction filter, and purge the flow channels. Replace the plunger seal. Check the plunge... Show Detail
- No:4272
- Date of Issue:2020/12/03 09:43
- Brached FAQ
-
No solvent is delivered at all
Click [PUMP]. [PUMP] is illuminated. Specify the flowrate. Click [CE] and eliminate the cause of the error. Show Detail
- No:4280
- Date of Issue:2020/12/03 10:20
- Brached FAQ
-
The solvent delivery pump runs, but does not pump solvent
Add mobile phase. Close the drain valve. Tighten the connection. If the leak persists after tightening the connection, then replace the connection. Cli... Show Detail
- No:4281
- Date of Issue:2020/12/03 10:22
- Brached FAQ
-
Solvent delivery is unstable and pump pulsation is large
Pump 2-propanol to rinse the check valve. However, if a buffer solution is being used, rinse the check valve with purified water first. Then rinse it by pump... Show Detail
- No:4282
- Date of Issue:2020/12/03 10:27
- Brached FAQ
-
Close the drain valve. Tighten the connection. If the leak persists after tightening the connection, then replace the connection. Click here for "Manual ... Show Detail
- No:4283
- Date of Issue:2020/12/03 10:29
- Brached FAQ
-
Ultrasonically clean or replace the line filter. Use the specified pipe. Inspect the column pressure and replace the column if clogged. Rinse the flow ... Show Detail
- No:4284
- Date of Issue:2020/12/03 10:30
- Brached FAQ
-
The rinse solution level for the seal automatic rinsing k...
Inspect (replace) the diaphragm. Tighten the connection. If the leak persists after tightening the connection, then replace the connection. Show Detail
- No:4285
- Date of Issue:2020/12/03 10:31
- Brached FAQ
51-60 of Total 72