• Font Size
  • S
  • M
  • L

FAQ within [ Solvent Delivery Module ]

51-60 of Total 72

6 / 8page
  • Error Code:[0x81D3] WARN : M-PHASE

    «Main Cause» The remaining volume of mobile phase is less than the alarm level. «Countermeasure» Replacement the mobile phase and reset the mobile phase qua... Show Detail

    • No:3641
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x81D2] WARN : M-PHASE

    «Main Cause» The remaining volume of mobile phase is less than the alarm level. «Countermeasure» Replacement the mobile phase and reset the mobile phase qua... Show Detail

    • No:3640
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x81D1] WARN : M-PHASE

    «Main Cause» The remaining volume of mobile phase is less than the alarm level. «Countermeasure» Replacement the mobile phase and reset the mobile phase qua... Show Detail

    • No:3639
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x81D0] WARN : M-PHASE

    «Main Cause» The remaining volume of mobile phase is less than the alarm level. «Countermeasure» Replacement the mobile phase and reset the mobile phase qua... Show Detail

    • No:3638
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0030] ERROR P.MAX

    «Main Cause» The discharge pressure of pump exceeds the Settings [P.Max]. Pumping will automatically stop. However, if "S-PROT" of auxiliary operation is Set... Show Detail

    • No:3637
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0034] ERROR LEAK

    «Main Cause» Detection of fluid leak due to pump X. «Countermeasure» Inspection the leak place of flow line, etc., and wipe off the fluid around the leak se... Show Detail

    • No:3636
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0032] ERROR HOME POS

    «Main Cause» Display when the home position of the pump motor cannot detection, the motor does not operate, or the motor is slipping. «Countermeasure» Once ... Show Detail

    • No:3635
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0031] ERROR P.MIN

    «Main Cause» Pump discharge pressure has fallen below the set min. pressure limit, [P.MIN]. Pumping will automatically stop. «Countermeasure» Flow line flui... Show Detail

    • No:3634
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x81E2] ERROR OVER HEAT

    «Main Cause» Interior temperature has risen to an abnormal level. «Countermeasure» Check to be sure that the internal fan can move, and that the rear exhaus... Show Detail

    • No:3633
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x81D4] OPEN DRAIN VALVE

    «Main Cause» Pump discharge pressure exceeded [P-PMAX] when purge. «Countermeasure» Open drain valve and press [Purge] button. If an error message appears d... Show Detail

    • No:3632
    • Date of Issue:2020/12/02 16:30

51-60 of Total 72

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.