      Error Code:[0x81E9] ERROR EXTERNAL INPUT «Main Cause» An error signal was input to the external input terminal. «Countermeasure» Check the connected external device and remove the cause of the error. Show Detail - No:3858
- Date of Issue:2020/12/02 16:43
 
- 
      Error Code:[0x81E8] ERROR NO CONST.PRS «Main Cause» The system controller specified the constant-pressure pump mode while a solvent delivery pump without constant-pressure pumping capability is be... Show Detail - No:3857
- Date of Issue:2020/12/02 16:43
 
- 
      Error Code:[0x81E0] ERROR NO LPGE MODE «Main Cause» The system controller specified the low-pressure gradient mode while a solvent delivery pump without low-pressure gradient capability is being u... Show Detail - No:3856
- Date of Issue:2020/12/02 16:43
 
- 
      Error Code:[0x5003] WARNING PUMP CONDITION «Main Cause» A sudden pressure change was detected during solvent delivery. «Countermeasure» There might be bubbles in the flow channel. Purge the mobile ph... Show Detail - No:3849
- Date of Issue:2020/12/02 16:43
 
- 
      Error Code:[0x5002] WARNING DEGAS PRESSURE «Main Cause» A seal problem occurred in the degassing unit vacuum flow channels. «Countermeasure» Check the vacuum pressure in the degassing unit. If the v... Show Detail - No:3848
- Date of Issue:2020/12/02 16:43
 
- 
      Error Code:[0x5001] WARNING F/W VERSION A-B «Main Cause» Different versions are being used by pumps A and B in the main system. «Countermeasure» The system can continue to be operated as it is, but fu... Show Detail - No:3847
- Date of Issue:2020/12/02 16:43
 
- 
      Error Code:[0x5000] FATAL CONNECTION A-B «Main Cause» Communication failed between pumps A and B in the main system. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or... Show Detail - No:3846
- Date of Issue:2020/12/02 16:43
 
- 
      
      «Main Cause» The pump was purged because pressure did not reach the lower pressure limit value, but an error occurred again, so solvent delivery was stopped.... Show Detail - No:3806
- Date of Issue:2020/12/02 16:43
 
- 
      
      «Main Cause» Solvent delivery was stopped because the pressure exceeded the system upper pressure limit value. «Countermeasure» Change the syste... Show Detail - No:3805
- Date of Issue:2020/12/02 16:43
- Last Updated:2020/12/04 18:40
 
- 
      Error Code:[0x81F1] WARN : DEGAS PRS «Main Cause» The abnormality of the vacuum pressure of the on-line degasser connect to this instrument will display for a certain time of time. Check the vac... Show Detail - No:3642
- Date of Issue:2020/12/02 16:30
 