Frequently Referenced FAQ

FAQ within [ Solvent Delivery Module ]

21-30 of Total 72

≪ 3 / 8page ≫
  • Error Code:[0x81D2] WARN : M-PHASE

    «Main Cause» The remaining volume of mobile phase is less than the alarm level. «Countermeasure» Replacement the mobile phase and reset the mobile phase qua... Show Detail

    • No:3640
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x81D3] WARN : M-PHASE

    «Main Cause» The remaining volume of mobile phase is less than the alarm level. «Countermeasure» Replacement the mobile phase and reset the mobile phase qua... Show Detail

    • No:3641
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x81F1] WARN : DEGAS PRS

    «Main Cause» The abnormality of the vacuum pressure of the Degassing Unit connect to this instrument will display for a certain time of time. Check the vacu... Show Detail

    • No:3642
    • Date of Issue:2020/12/02 16:30
    • Last Updated:2026/11/04 13:31
  • Error Code:[0x8003]

    «Main Cause» Solvent delivery was stopped because the pressure exceeded the system upper pressure limit value. «Countermeasure» Change the syste... Show Detail

    • No:3805
    • Date of Issue:2020/12/02 16:43
    • Last Updated:2020/12/04 18:40
  • Error Code:[0x8010]

    «Main Cause» The pump was purged because pressure did not reach the lower pressure limit value, but an error occurred again, so solvent delivery was stopped.... Show Detail

    • No:3806
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x5000] FATAL CONNECTION A-B

    «Main Cause» Communication failed between pumps A and B in the main system. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or... Show Detail

    • No:3846
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x5001] WARNING F/W VERSION A-B

    «Main Cause» Different versions are being used by pumps A and B in the main system. «Countermeasure» The system can continue to be operated as it is, but fu... Show Detail

    • No:3847
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x5002] WARNING DEGAS PRESSURE

    «Main Cause» A seal problem occurred in the degassing unit vacuum flow channels. «Countermeasure» Check the vacuum pressure in the degassing unit. If the v... Show Detail

    • No:3848
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x5003] WARNING PUMP CONDITION

    «Main Cause» A sudden pressure change was detected during solvent delivery. «Countermeasure» There might be bubbles in the flow channel. Purge the mobile ph... Show Detail

    • No:3849
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x81E0] ERROR NO LPGE MODE

    «Main Cause» The system controller specified the low-pressure gradient mode while a solvent delivery pump without low-pressure gradient capability is being u... Show Detail

    • No:3856
    • Date of Issue:2020/12/02 16:43