(HPLC) Abnormal pressure value of Solvent Delivery Module
Zero adjustment of pressure sensor. 1. Set to [ZERO ADJUST] in the [CONTROL] group. * For LC-20A / 30A, [ZERO ADJ] 2. Open the drain valve to remove ... Show Detail
- No:4768
- Date of Issue:2021/01/13 14:16
-
(HPLC) Replace of Mixer 0.5-2.6mL, 100uL line filter
If the back pressure is more than 1MPa when the flowrate is 1mL/min and the mixer outlet is clogged. Remove the male nut and loosen the screw to replace th... Show Detail
- No:4537
- Date of Issue:2020/12/11 16:25
-
Replacement of Plunger Seal (HPLC)
Plunger seal parts number: Standard: 228-52711-93 Plunger seal UHP(with back up ring) Option: 228-52711-94 Plunger seal, back upring, GFP *The m... Show Detail
- No:4413
- Date of Issue:2020/12/04 16:01
- Last Updated:2021/07/12 15:58
- Brached FAQ
-
Maintenance for LC-40 (Pump) series (Video)
Consumable Part No. 228-57501-92 Line Filter (LC-40D X3/XS, 40B X3) 228-35871-96 Line Filter (LC-40D/XR, 40B XR) 228-70796-42 Line Filter (LC-40D ... Show Detail
- No:4340
- Date of Issue:2020/12/03 13:55
- Brached FAQ
-
Check whether the noise disappears when solvent delivery stops. Check the solvent delivery flow status (pulsation). If pulsation is large, rinse the flow ch... Show Detail
- No:4300
- Date of Issue:2020/12/03 11:06
- Brached FAQ
-
Check whether the noise disappears when solvent delivery stops. Check the solvent delivery flow status (pulsation). If pulsation is large, rinse the flow ch... Show Detail
- No:4299
- Date of Issue:2020/12/03 11:05
- Brached FAQ
-
The baseline undulates with the pump stroke
Click [PURGE] to purge the bubbles. Alternatively, connect a syringe to the drain tube connection port to suction out the bubbles. Click here for Manual P... Show Detail
- No:4296
- Date of Issue:2020/12/03 11:00
- Brached FAQ
-
Inspect the suction filter for blockage. Also rinse them as described below. 1) Flush with 17 % aqueous phosphate solution for about 30 minutes at a flowr... Show Detail
- No:4294
- Date of Issue:2020/12/03 10:57
- Brached FAQ
-
Inspect the suction filter for blockage.Also rinse them as described below. 1) Flush with 17 % aqueous phosphate solution for about 30 minutes at a flowra... Show Detail
- No:4293
- Date of Issue:2020/12/03 10:52
- Brached FAQ