  • (LC) Consumables (Lamp for Detector)

    Appearance Appearance Show Detail

    • No:6579
    • Date of Issue:2020/08/25 09:38
    • Brached FAQ
  • (LC) Consumables (Flow Cell for Detector)

    Appearance Appearance Appearance ※Light blue areas are Field Engineer replacement service parts and contact... Show Detail

    • No:6584
    • Date of Issue:2020/08/25 11:27
    • Brached FAQ
  • Maintenance for i-Series(LC-2030/40/50/60) (Video)

    Click here for (LC) Consumables (Suction Filter (Pump)) Click here for (LC) Consumables (Check Valve) Click here for(LC... Show Detail

    • No:4341
    • Date of Issue:2020/08/25 17:34
    • Brached FAQ
  • How to set the exposure time and how to reset the total l...

    1. If the LabSolutions analysis program is running, exit the analysis program. 2. Click the (Start) menu to display the list of applications and then click... Show Detail

    • No:3015
    • Date of Issue:2020/11/09 15:36
    • Last Updated:2025/12/03 13:45
  • Cleaning/Replacing the ELSD Nebulizer

    CAUTION: Take sufficient care while handling the nebulizer • Never touch the IN tubing fitting section (B) of the nebulizer. • Replace the nebulizer if it ... Show Detail

    • No:3072
    • Date of Issue:2020/11/11 09:50
    • Last Updated:2023/10/18 10:52
  • Cleaning the Interior of the ELSD

    1. Set the temperature to 100°C and flow gas at 350kPa. 2.Pump a suitable mobile phase at 2mL/min. CAUTION: Select a mobile phase in which the analyzed s... Show Detail

    • No:3073
    • Date of Issue:2020/11/11 09:54
    • Last Updated:2021/01/07 16:32
  • Initial IP address, Log in ID, Pass word (HPLC)

    · Instruments [CALIBRATION] menu PW: 00000 · CBM-20(lite)/40(lite), SCL-40 IP address: 192.168.200.99 Sub-net mask: 255.255.255.0 Web controller: http... Show Detail

    • No:4395
    • Date of Issue:2020/11/11 10:50
    • Last Updated:2026/02/13 11:31
  • Error Code:[0x9304]

    «Main Cause» There is a fault in the non-volatile memory of the detector or a non-volatile memory write error has occurred. «Countermeasure» Perform calibra... Show Detail

    • No:3403
    • Date of Issue:2020/12/02 16:09
  • Error Code:[0x9321]

    «Main Cause» A liquid leak has occurred in the detector. «Countermeasure» Inspect the detector for liquid leaks, and wipe away any liquids from around the l... Show Detail

    • No:3404
    • Date of Issue:2020/12/02 16:09
  • Error Code:[0x9341]

    «Main Cause» The buffer memory used for transmitting data from the instrument to the workstation is full. «Countermeasure» Check the communication environm... Show Detail

    • No:3405
    • Date of Issue:2020/12/02 16:09