  • Error Code:[0x58E6] FATAL D2 LAMP NOT LIT

    «Main Cause» The D2 lamp failed to illuminate. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of Lamp (HPLC) Show Detail

    • No:4082
    • Date of Issue:2020/12/02 16:45
    • Last Updated:2020/12/07 11:13
  • Error Code:[0x58E5] ERROR D2 NOT CONNECTED

    «Main Cause» D2 lamp is not connected. «Countermeasure» Connect the D2 lamp. Show Detail

    • No:4081
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x58E4] FATAL D2 LAMP DRIVER

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:4080
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x58DC] FATAL DETECTOR

    «Main Cause» Initialization of internal actions failed. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representat... Show Detail

    • No:4079
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x58DA] FATAL DETECTOR

    «Main Cause» Initialization of internal actions failed. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representat... Show Detail

    • No:4078
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x58D6] FATAL INNER FAN

    «Main Cause» A rotation problem was detected for the internal fan. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:4077
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x58D5] ERROR D2 LAMP NOT LIT

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:4076
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x58D4] ERROR D2 LAMP NOT LIT

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:4075
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x58D3] ERROR D2 LAMP NOT LIT

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:4074
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x58D1] FATAL FILTER POSITION

    «Main Cause» Detection of the filter motor home position failed. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service re... Show Detail

    • No:4073
    • Date of Issue:2020/12/02 16:45