FAQ - Frequently Asked Questions
- Products Search > HPLC > Detector
Keyword Search
Frequently Referenced FAQ
FAQ within [ Detector ]
-
«Main Cause» A broken circuit may have occurred in the lamp housing heater. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or... Show Detail
- No:4112
- Date of Issue:2020/12/02 16:46
-
«Main Cause» A broken circuit may have occurred in the cell heater. «Countermeasure» If the error is not resolved after reconnecting the cell, switch the in... Show Detail
- No:4111
- Date of Issue:2020/12/02 16:46
-
«Main Cause» A broken circuit may have occurred in the monochromator heater. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales o... Show Detail
- No:4110
- Date of Issue:2020/12/02 16:46
-
«Main Cause» A broken circuit may have occurred in the lamp housing sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or... Show Detail
- No:4109
- Date of Issue:2020/12/02 16:46
-
«Main Cause» A broken circuit may have occurred in the leak sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service... Show Detail
- No:4108
- Date of Issue:2020/12/02 16:46
-
«Main Cause» A leak sensor short circuit may have occurred. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service represe... Show Detail
- No:4107
- Date of Issue:2020/12/02 16:45
-
«Main Cause» A broken circuit may have occurred in the cell temperature sensor. «Countermeasure» If the error is not resolved after reconnecting the cell, s... Show Detail
- No:4106
- Date of Issue:2020/12/02 16:45
-
«Main Cause» Cannot write data to D2 memory. «Countermeasure» Check the D2 lamp connection. If that does not resolved the error, replace the D2 ... Show Detail
- No:4105
- Date of Issue:2020/12/02 16:45
- Last Updated:2020/12/07 11:14
-
«Main Cause» Information in the D2 lamp memory could not be read. «Countermeasure» Contact a Shimadzu sales or service representative. Show Detail
- No:4104
- Date of Issue:2020/12/02 16:45
-
«Main Cause» Analysis was started without registering the cell in the main unit. «Countermeasure» Register the cell in the main unit using Utility. Show Detail
- No:4103
- Date of Issue:2020/12/02 16:45
51-60 of Total 266