«Main Cause» Error occurred during system check. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3645
- Date of Issue:2020/12/02 16:31
-
«Main Cause» There is a PC data buffer overflow. «Countermeasure» Check the PC status and the cable connections. If the error persists, turn the power OFF ... Show Detail
- No:3646
- Date of Issue:2020/12/02 16:31
-
«Main Cause» Hard disk initialization failed. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3647
- Date of Issue:2020/12/02 16:31
-
«Main Cause» Hard disk is full. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3648
- Date of Issue:2020/12/02 16:31
-
«Main Cause» Detector buffer memory overflowed. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3649
- Date of Issue:2020/12/02 16:31
-
«Main Cause» The signal is saturated because the exposure time for the PDA elements is inappropriate. «Countermeasure» Perform exposure time optimization u... Show Detail
- No:3650
- Date of Issue:2020/12/02 16:31
-
«Main Cause» Ethernet communication with the PC timed out. «Countermeasure» Check the power supply of the instrument and cable connections. Restart the PDA... Show Detail
- No:3651
- Date of Issue:2020/12/02 16:31
-
«Main Cause» Ethernet communication with the CBM timed out. «Countermeasure» Check the network environment. Turn the power of PDA detector and CBM OFF once... Show Detail
- No:3652
- Date of Issue:2020/12/02 16:31
-
«Main Cause» A device with the same IP address as the instrument exists on the same network. «Countermeasure» Change the IP address of the instrument or de... Show Detail
- No:3653
- Date of Issue:2020/12/02 16:31
-
«Main Cause» Temperature-controlled flow cell malfunctioned, control circuit malfunctioned. «Countermeasure» Turn the power OFF and contact a Shimadzu sale... Show Detail
- No:3654
- Date of Issue:2020/12/02 16:31