FAQ - Frequently Asked Questions
- Products Search > HPLC > Detector
Keyword Search
Frequently Referenced FAQ
FAQ within [ Detector ]
- 
      
      «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail - No:4122
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:04
 
- 
      
      «Main Cause» The W lamp or W lamp illumination circuit is not functioning properly. «Countermeasure» Replace the W lamp. Click here for Repla... Show Detail - No:4121
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:03
 
- 
      
      «Main Cause» The lamp housing became hot. «Countermeasure» Implement corrective measures in the following sequence. (1) Check whether the fan on the back of... Show Detail - No:4120
- Date of Issue:2020/12/02 16:46
 
- 
      
      «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail - No:4119
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:02
 
- 
      
      «Main Cause» The lamp housing opened or the cover sensor is not functioning properly. «Countermeasure» Close the lamp cover. If that does not resolve the er... Show Detail - No:4118
- Date of Issue:2020/12/02 16:46
 
- 
      
      «Main Cause» The W lamp failed to illuminate due to a broken filament or the illumination circuit is not functioning properly. «Countermeasure» ... Show Detail - No:4117
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:02
 
- 
      
      «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail - No:4116
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 10:59
 
- 
      
      «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail - No:4115
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:01
 
- 
      
      «Main Cause» A rotation problem was detected for the fan on the back of the instrument. «Countermeasure» Switch the instrument power OFF and contact a Shima... Show Detail - No:4114
- Date of Issue:2020/12/02 16:46
 
- 
      
      «Main Cause» The calibration information for Undefined slots is the default value. «Countermeasure» Register the calibration information using PDAUtility. Show Detail - No:4113
- Date of Issue:2020/12/02 16:46
 
41-50 of Total 266