  • Error Code:[0x93DA]

    «Main Cause» The instrument is turned off because the ADC module on the board is malfunctioning. The instrument is stopped. «Countermeasure» Press the (pow... Show Detail

    • No:3436
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93DB]

    «Main Cause» The optical unit temperature does not reach the set temperature. Temperature control is turned off. «Countermeasure» Contact a Shimadzu sales o... Show Detail

    • No:3437
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93DC]

    «Main Cause» The heater of the spectrmeter is short or there is a fault in the circuit. «Countermeasure» Press the (power switch) to turn the power OFF and ... Show Detail

    • No:3438
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93E2]

    «Main Cause» Failed to receive the detector signal. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:3439
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93E4]

    «Main Cause» Failed to receive the detector signal. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:3440
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93F1]

    «Main Cause» Could not detect the filter home position in the detector. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shima... Show Detail

    • No:3441
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93F2]

    «Main Cause» Could not detect the grating home position in the detector. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shim... Show Detail

    • No:3442
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93F4]

    «Main Cause» Results of wavelength accuracy check at instrument startup differ by more than 1nm. There may be bubbles in the flow cell. «Countermeasure» Per... Show Detail

    • No:3443
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x8904]

    «Main Cause» Invalid EEPROM parameter. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3643
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8921]

    «Main Cause» Mobile phase has leaked inside the instrument. «Countermeasure» Check the plumbing. Show Detail

    • No:3644
    • Date of Issue:2020/12/02 16:30