  • Error Code:[0x5ACA]

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail

    • No:4122
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:04
  • Error Code:[0x5AC9]

    «Main Cause» The W lamp or W lamp illumination circuit is not functioning properly. «Countermeasure» Replace the W lamp. Click here for Repla... Show Detail

    • No:4121
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:03
  • Error Code:[0x5AC8]

    «Main Cause» The lamp housing became hot. «Countermeasure» Implement corrective measures in the following sequence. (1) Check whether the fan on the back of... Show Detail

    • No:4120
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AC7]

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail

    • No:4119
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:02
  • Error Code:[0x5AC6]

    «Main Cause» The lamp housing opened or the cover sensor is not functioning properly. «Countermeasure» Close the lamp cover. If that does not resolve the er... Show Detail

    • No:4118
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AC4]

    «Main Cause» The W lamp failed to illuminate due to a broken filament or the illumination circuit is not functioning properly. «Countermeasure» ... Show Detail

    • No:4117
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:02
  • Error Code:[0x5AC3]

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail

    • No:4116
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 10:59
  • Error Code:[0x5AC2]

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail

    • No:4115
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:01
  • Error Code:[0x5A85]

    «Main Cause» A rotation problem was detected for the fan on the back of the instrument. «Countermeasure» Switch the instrument power OFF and contact a Shima... Show Detail

    • No:4114
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5A84]

    «Main Cause» The calibration information for Undefined slots is the default value. «Countermeasure» Register the calibration information using PDAUtility. Show Detail

    • No:4113
    • Date of Issue:2020/12/02 16:46

41-50 of Total 266