  • Error Code:[0x5ADA]

    «Main Cause» PDA detector initialization failed. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:4132
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AD9]

    «Main Cause» A broken circuit may have occurred in the monochromator temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shim... Show Detail

    • No:4131
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AD8]

    «Main Cause» The monochromator became hot. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:4130
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AD6]

    «Main Cause» A rotation problem was detected for the internal fan. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:4129
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AD5]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail

    • No:4128
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:06
  • Error Code:[0x5AD4]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail

    • No:4127
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:06
  • Error Code:[0x5AD3]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail

    • No:4126
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:05
  • Error Code:[0x5AD1]

    «Main Cause» Detector initialization failed. «Countermeasure» Implement corrective measures in the following sequence. (1) Reconfirm that the flow cell is i... Show Detail

    • No:4125
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5ACD]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:4124
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5ACC]

    «Main Cause» The W lamp is not connected or the illumination circuit is not functioning properly. «Countermeasure» Implement corrective measures... Show Detail

    • No:4123
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:04