• Font Size
  • S
  • M
  • L

FAQ within [ Detector ]

31-40 of Total 266

4 / 27page
  • Error Code:[0x93CC]

    «Main Cause» The W lamp connector is not connected. Alternatively, the W lamp cable is broken or there is a fault in the board. «Countermeasure» ... Show Detail

    • No:3426
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2020/12/07 11:19
  • Error Code:[0x93CD]

    «Main Cause» There is a fault in the lighting circuit. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales or servi... Show Detail

    • No:3427
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93D1]

    «Main Cause» Could not detect the filter home position in the detector. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shima... Show Detail

    • No:3428
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93D2]

    «Main Cause» The temperature of the detector board is high. «Countermeasure» Follow the procedures below in the following order. (1) Check whether the fan ... Show Detail

    • No:3429
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93D3]

    «Main Cause» The fan and the inspiratory port are closed or a fault in the lighting circuit. «Countermeasure» Follow the procedures below in the following o... Show Detail

    • No:3430
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93D4]

    «Main Cause» There is a fault in the lighting circuit. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales or servi... Show Detail

    • No:3431
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93D5]

    «Main Cause» There is a fault in the lighting circuit. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales or servi... Show Detail

    • No:3432
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93D6]

    «Main Cause» The lamp is turned off because a rotation error was detected in the detector fan. «Countermeasure» Press the (power switch) to turn the power O... Show Detail

    • No:3433
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93D8]

    «Main Cause» Temperature control is turned off because the temperature of the optical unit is high. «Countermeasure» Press the (power switch) to turn the po... Show Detail

    • No:3434
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93D9]

    «Main Cause» Temperature control is turned off because the temperature sensor of the optical unit is malfunctioning. «Countermeasure» Press the (power switc... Show Detail

    • No:3435
    • Date of Issue:2020/12/02 16:10

31-40 of Total 266

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.