• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Detector ]

21-30 of Total 266

3 / 27page
  • Error Code:[0x5AF2]

    «Main Cause» A flow cell not compatible with the instrument was connected. «Countermeasure» Connect a flow cell for PDA detectors. Show Detail

    • No:4142
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AF1]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail

    • No:4141
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:08
  • Error Code:[0x5AF0]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail

    • No:4140
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:07
  • Error Code:[0x5AE9]

    «Main Cause» The cell temperature could not be controlled to the temperature specified in the method because a cell that does not support temperature-control... Show Detail

    • No:4139
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AE8]

    «Main Cause» Communication failed between the operation panel and the instrument. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sa... Show Detail

    • No:4138
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AE7]

    «Main Cause» Cannot communicate normally between the instrument and workstation. «Countermeasure» Check the communication environment and operating status b... Show Detail

    • No:4137
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AE6]

    «Main Cause» The D2 lamp failed to illuminate. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of Lamp (HPLC) Show Detail

    • No:4136
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:07
  • Error Code:[0x5AE5]

    «Main Cause» D2 lamp is not connected. «Countermeasure» Connect the D2 lamp. Show Detail

    • No:4135
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AE4]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:4134
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AE2]

    «Main Cause» Detector data processing did not function properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service re... Show Detail

    • No:4133
    • Date of Issue:2020/12/02 16:46

21-30 of Total 266

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.