Frequently Referenced FAQ

FAQ within [ Detector ]

21-30 of Total 266

≪ 3 / 27page ≫
  • Error Code:[0x937A]

    «Main Cause» The lamp is turned off because the temperature sensor of the detector lamp housing is malfunctioning. «Countermeasure» Press the (power switc... Show Detail

    • No:3416
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93C2]

    «Main Cause» There is a fault in the D2 lamp or a fault in the lighting circuit. «Countermeasure» Replace the D2 lamp. Click here for Replace... Show Detail

    • No:3417
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2020/12/07 11:15
  • Error Code:[0x93C3]

    «Main Cause» A lighting defect has occurred due to a broken filament in the D2 lamp or a fault in the constant current circuit, a lighting malfunction has... Show Detail

    • No:3418
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2020/12/07 11:15
  • Error Code:[0x93C4]

    «Main Cause» This is a lighting defect caused by a broken filament in the W lamp, or due to a fault in the lighting circuit. «Countermeasure» Re... Show Detail

    • No:3419
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2020/12/07 11:16
  • Error Code:[0x93C6]

    «Main Cause» The lamp house cover is open or there is a fault in the cover sensor. «Countermeasure» Close the lamp house cover. If the error is not resolve... Show Detail

    • No:3420
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93C7]

    «Main Cause» There is a fault in the D2 lamp or a fault in the lighting circuit. «Countermeasure» Replace the D2 lamp. Click here for Replace... Show Detail

    • No:3421
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2020/12/07 11:16
  • Error Code:[0x93C8]

    «Main Cause» The lamp is turned off because the temperature of the detector lamp housing is high. «Countermeasure» Follow the procedures below in the foll... Show Detail

    • No:3422
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93C9]

    «Main Cause» There is a fault in the W lamp lighting circuit. «Countermeasure» Replace the W lamp. Click here for Replacement method of Lamp ... Show Detail

    • No:3423
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2020/12/07 11:17
  • Error Code:[0x93CA]

    «Main Cause» There is a fault in the D2 lamp or a fault in the lighting circuit. «Countermeasure» Replace the D2 lamp. Click here for Replace... Show Detail

    • No:3424
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2020/12/07 11:17
  • Error Code:[0x93CB]

    «Main Cause» The D2 lamp connector is not connected. Alternatively, the D2 lamp cable is broken or there is a fault in the board. «Countermeasure»... Show Detail

    • No:3425
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2020/12/07 11:18