FAQ - Frequently Asked Questions
- Products Search > HPLC > Detector
Keyword Search
Frequently Referenced FAQ
FAQ within [ Detector ]
-
«Main Cause» There is a fault in the control circuit of the temperature control cell. «Countermeasure» Replace the temperature control cell. Show Detail
- No:3411
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The IP address of the instrument is duplicated on a different device. «Countermeasure» Change the IP addresses so there are no duplicates. If ... Show Detail
- No:3410
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The detector cell temperature does not reach the set temperature. «Countermeasure» Replace the temperature control cell. If the error is not ... Show Detail
- No:3409
- Date of Issue:2020/12/02 16:09
-
«Main Cause» Communication between the instrument and the workstation was interrupted. «Countermeasure» Check the communication environment and operating st... Show Detail
- No:3408
- Date of Issue:2020/12/02 16:09
-
«Main Cause» Light intensity was exceeded because the detector took a long exposure time. «Countermeasure» Calibrate the exposure time. Show Detail
- No:3407
- Date of Issue:2020/12/02 16:09
-
«Main Cause» Recovery failed when resending transmission data from the instrument to the workstation «Countermeasure» Check the communication environment a... Show Detail
- No:3406
- Date of Issue:2020/12/02 16:09
-
«Main Cause» The buffer memory used for transmitting data from the instrument to the workstation is full. «Countermeasure» Check the communication environm... Show Detail
- No:3405
- Date of Issue:2020/12/02 16:09
-
«Main Cause» A liquid leak has occurred in the detector. «Countermeasure» Inspect the detector for liquid leaks, and wipe away any liquids from around the l... Show Detail
- No:3404
- Date of Issue:2020/12/02 16:09
-
«Main Cause» There is a fault in the non-volatile memory of the detector or a non-volatile memory write error has occurred. «Countermeasure» Perform calibra... Show Detail
- No:3403
- Date of Issue:2020/12/02 16:09
-
Initial IP address, Log in ID, Pass word (HPLC)
· Instruments [CALIBRATION] menu PW: 00000 · CBM-20(lite)/40(lite), SCL-40 IP address: 192.168.200.99 Sub-net mask: 255.255.255.0 Web contro... Show Detail
- No:4395
- Date of Issue:2020/11/11 10:50
251-260 of Total 266