«Main Cause» There is a fault in the D2 lamp or a fault in the lighting circuit. «Countermeasure» Replace the D2 lamp. Click here for Replace... Show Detail
- No:3421
- Date of Issue:2020/12/02 16:10
- Last Updated:2020/12/07 11:16
-
«Main Cause» The lamp house cover is open or there is a fault in the cover sensor. «Countermeasure» Close the lamp house cover. If the error is not resolve... Show Detail
- No:3420
- Date of Issue:2020/12/02 16:10
-
«Main Cause» This is a lighting defect caused by a broken filament in the W lamp, or due to a fault in the lighting circuit. «Countermeasure» Re... Show Detail
- No:3419
- Date of Issue:2020/12/02 16:10
- Last Updated:2020/12/07 11:16
-
«Main Cause» A lighting defect has occurred due to a broken filament in the D2 lamp or a fault in the constant current circuit, a lighting malfunction has... Show Detail
- No:3418
- Date of Issue:2020/12/02 16:10
- Last Updated:2020/12/07 11:15
-
«Main Cause» There is a fault in the D2 lamp or a fault in the lighting circuit. «Countermeasure» Replace the D2 lamp. Click here for Replace... Show Detail
- No:3417
- Date of Issue:2020/12/02 16:10
- Last Updated:2020/12/07 11:15
-
«Main Cause» The lamp is turned off because the temperature sensor of the detector lamp housing is malfunctioning. «Countermeasure» Press the (power switc... Show Detail
- No:3416
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The leak sensor of the detector is malfunctioning. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sale... Show Detail
- No:3415
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The leak sensor of the detector is malfunctioning. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales... Show Detail
- No:3414
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The [Cell Temperature] settings in the method parameters are set to a value outside the setting range. «Countermeasure» Set the [Cell Temperatu... Show Detail
- No:3413
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The temperature control cell temperature does not lower to the set temperature even after 5 minutes have elapsed. «Countermeasure» Set the [Cel... Show Detail
- No:3412
- Date of Issue:2020/12/02 16:10