FAQ within [ Detector ]

231-240 of Total 266

24 / 27page
  • Error Code:[0x5AD4]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail

    • No:4127
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:06
  • Error Code:[0x5AD5]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail

    • No:4128
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:06
  • Error Code:[0x5AD6]

    «Main Cause» A rotation problem was detected for the internal fan. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:4129
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AD8]

    «Main Cause» The monochromator became hot. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:4130
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AD9]

    «Main Cause» A broken circuit may have occurred in the monochromator temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shim... Show Detail

    • No:4131
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5ADA]

    «Main Cause» PDA detector initialization failed. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:4132
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AE2]

    «Main Cause» Detector data processing did not function properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service re... Show Detail

    • No:4133
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AE4]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:4134
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AE5]

    «Main Cause» D2 lamp is not connected. «Countermeasure» Connect the D2 lamp. Show Detail

    • No:4135
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AE6]

    «Main Cause» The D2 lamp failed to illuminate. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of Lamp (HPLC) Show Detail

    • No:4136
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:07