«Main Cause» The W lamp failed to illuminate due to a broken filament or the illumination circuit is not functioning properly. «Countermeasure» ... Show Detail
- No:4117
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:02
-
«Main Cause» The lamp housing opened or the cover sensor is not functioning properly. «Countermeasure» Close the lamp cover. If that does not resolve the er... Show Detail
- No:4118
- Date of Issue:2020/12/02 16:46
-
«Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail
- No:4119
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:02
-
«Main Cause» The lamp housing became hot. «Countermeasure» Implement corrective measures in the following sequence. (1) Check whether the fan on the back of... Show Detail
- No:4120
- Date of Issue:2020/12/02 16:46
-
«Main Cause» The W lamp or W lamp illumination circuit is not functioning properly. «Countermeasure» Replace the W lamp. Click here for Repla... Show Detail
- No:4121
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:03
-
«Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail
- No:4122
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:04
-
«Main Cause» The W lamp is not connected or the illumination circuit is not functioning properly. «Countermeasure» Implement corrective measures... Show Detail
- No:4123
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:04
-
«Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail
- No:4124
- Date of Issue:2020/12/02 16:46
-
«Main Cause» Detector initialization failed. «Countermeasure» Implement corrective measures in the following sequence. (1) Reconfirm that the flow cell is i... Show Detail
- No:4125
- Date of Issue:2020/12/02 16:46
-
«Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail
- No:4126
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:05