FAQ within [ Detector ]

221-230 of Total 266

23 / 27page
  • Error Code:[0x5AC4]

    «Main Cause» The W lamp failed to illuminate due to a broken filament or the illumination circuit is not functioning properly. «Countermeasure» ... Show Detail

    • No:4117
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:02
  • Error Code:[0x5AC6]

    «Main Cause» The lamp housing opened or the cover sensor is not functioning properly. «Countermeasure» Close the lamp cover. If that does not resolve the er... Show Detail

    • No:4118
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AC7]

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail

    • No:4119
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:02
  • Error Code:[0x5AC8]

    «Main Cause» The lamp housing became hot. «Countermeasure» Implement corrective measures in the following sequence. (1) Check whether the fan on the back of... Show Detail

    • No:4120
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AC9]

    «Main Cause» The W lamp or W lamp illumination circuit is not functioning properly. «Countermeasure» Replace the W lamp. Click here for Repla... Show Detail

    • No:4121
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:03
  • Error Code:[0x5ACA]

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail

    • No:4122
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:04
  • Error Code:[0x5ACC]

    «Main Cause» The W lamp is not connected or the illumination circuit is not functioning properly. «Countermeasure» Implement corrective measures... Show Detail

    • No:4123
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:04
  • Error Code:[0x5ACD]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:4124
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AD1]

    «Main Cause» Detector initialization failed. «Countermeasure» Implement corrective measures in the following sequence. (1) Reconfirm that the flow cell is i... Show Detail

    • No:4125
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5AD3]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail

    • No:4126
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:05