FAQ within [ Detector ]

211-220 of Total 266

≪ 22 / 27page ≫
  • Error Code:[0x8945]

    «Main Cause» The signal is saturated because the exposure time for the PDA elements is inappropriate. «Countermeasure» Perform exposure time optimization u... Show Detail

    • No:3650
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8944]

    «Main Cause» Detector buffer memory overflowed. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3649
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8943]

    «Main Cause» Hard disk is full. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3648
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8942]

    «Main Cause» Hard disk initialization failed. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3647
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8941]

    «Main Cause» There is a PC data buffer overflow. «Countermeasure» Check the PC status and the cable connections. If the error persists, turn the power OFF ... Show Detail

    • No:3646
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8922]

    «Main Cause» Error occurred during system check. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3645
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8921]

    «Main Cause» Mobile phase has leaked inside the instrument. «Countermeasure» Check the plumbing. Show Detail

    • No:3644
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8904]

    «Main Cause» Invalid EEPROM parameter. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3643
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x93F4]

    «Main Cause» Results of wavelength accuracy check at instrument startup differ by more than 1nm. There may be bubbles in the flow cell. «Countermeasure» Per... Show Detail

    • No:3443
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x93F2]

    «Main Cause» Could not detect the grating home position in the detector. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shim... Show Detail

    • No:3442
    • Date of Issue:2020/12/02 16:10