• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Detector ]

201-210 of Total 266

21 / 27page
  • Error Code:[0x8978]

    «Main Cause» Leak sensor is short. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3660
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8967]

    «Main Cause» The wavelength of the installed cell is not calibrated. «Countermeasure» Perform wavelength calibration. Show Detail

    • No:3659
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8966]

    «Main Cause» The exposure time of the installed cell is not adjusted. «Countermeasure» Set the exposure time. Show Detail

    • No:3658
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8965]

    «Main Cause» (1) An unregistered cell was installed in the instrument. (2) Failed to read cell ID. «Countermeasure» (1) Perform cell registration, exposure... Show Detail

    • No:3657
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8964]

    «Main Cause» Cell is not installed or the cell cable is not connected. «Countermeasure» Install the cell and connect the cable correctly. Show Detail

    • No:3656
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8962]

    «Main Cause» Temperature-controlled flow cell malfunctioned, control circuit malfunctioned. «Countermeasure» Turn the power OFF and contact a Shimadzu sales... Show Detail

    • No:3655
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8961]

    «Main Cause» Temperature-controlled flow cell malfunctioned, control circuit malfunctioned. «Countermeasure» Turn the power OFF and contact a Shimadzu sale... Show Detail

    • No:3654
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8956]

    «Main Cause» A device with the same IP address as the instrument exists on the same network. «Countermeasure» Change the IP address of the instrument or de... Show Detail

    • No:3653
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8954]

    «Main Cause» Ethernet communication with the CBM timed out. «Countermeasure» Check the network environment. Turn the power of PDA detector and CBM OFF once... Show Detail

    • No:3652
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8952]

    «Main Cause» Ethernet communication with the PC timed out. «Countermeasure» Check the power supply of the instrument and cable connections. Restart the PDA... Show Detail

    • No:3651
    • Date of Issue:2020/12/02 16:31

201-210 of Total 266

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.