FAQ within [ Detector ]
-
«Main Cause» Communication between the instrument and system controller was interrupted. «Countermeasure» · Check whether PDA detector and system controller... Show Detail
- No:4097
- Date of Issue:2020/12/02 16:45
-
«Main Cause» An instrument IP address conflict occurred with another device. «Countermeasure» Resolve the IP address conflict. If that does not resolve the ... Show Detail
- No:4098
- Date of Issue:2020/12/02 16:45
-
«Main Cause» The analysis start time received from the system controller is later than the current PDA detector time. «Countermeasure» Restart the PDA detec... Show Detail
- No:4099
- Date of Issue:2020/12/02 16:45
-
«Main Cause» The temperature-controlled cell control circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shi... Show Detail
- No:4100
- Date of Issue:2020/12/02 16:45
-
«Main Cause» The temperature did not decrease to the setting temperature even after five minutes of temperature control. «Countermeasure» Specify a somewhat... Show Detail
- No:4101
- Date of Issue:2020/12/02 16:45
-
«Main Cause» Analysis was started without connecting the flow cell. «Countermeasure» Connect the cell to the main unit. Show Detail
- No:4102
- Date of Issue:2020/12/02 16:45
-
«Main Cause» Analysis was started without registering the cell in the main unit. «Countermeasure» Register the cell in the main unit using Utility. Show Detail
- No:4103
- Date of Issue:2020/12/02 16:45
-
«Main Cause» Information in the D2 lamp memory could not be read. «Countermeasure» Contact a Shimadzu sales or service representative. Show Detail
- No:4104
- Date of Issue:2020/12/02 16:45
-
«Main Cause» Cannot write data to D2 memory. «Countermeasure» Check the D2 lamp connection. If that does not resolved the error, replace the D2 ... Show Detail
- No:4105
- Date of Issue:2020/12/02 16:45
- Last Updated:2020/12/07 11:14
-
«Main Cause» A broken circuit may have occurred in the cell temperature sensor. «Countermeasure» If the error is not resolved after reconnecting the cell, s... Show Detail
- No:4106
- Date of Issue:2020/12/02 16:45