Frequently Referenced FAQ

FAQ within [ Detector ]

191-200 of Total 266

≪ 20 / 27page ≫
  • Error Code:[0x89CA]

    «Main Cause» A probable cause is that the D2 lamp heater circuit malfunctioned. «Countermeasure» Click here for Replacement method of Lamp (HPLC... Show Detail

    • No:3670
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:21
  • Error Code:[0x89C9]

    «Main Cause» A probable cause is that the W lamp malfunctioned. «Countermeasure» Check the total operating time of the lamp, and replace the lamp. If this ... Show Detail

    • No:3669
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x89C8]

    «Main Cause» A probable cause is that the cooling fan blast flow is poor, or the cooling fan or heater circuit malfunctioned. «Countermeasure» Ensure at le... Show Detail

    • No:3668
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x89C7]

    «Main Cause» A probable cause is that the D2 lamp heater circuit malfunctioned. «Countermeasure» Check the total operating time of the lamp, and replace th... Show Detail

    • No:3667
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x89C6]

    «Main Cause» The lamp cover was not in place at instrument startup. «Countermeasure» Mount the lamp cover, and turn the lamp on again. Show Detail

    • No:3666
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x89C4]

    «Main Cause» Probable causes are the W lamp connector is not connected, or the W lamp malfunctioned. «Countermeasure» Check the W lamp connections. Check t... Show Detail

    • No:3665
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x89C3]

    «Main Cause» Probable causes are the D2 lamp connector is not connected, the D2 lamp malfunctioned, or the lighting circuit malfunctioned. «Counter... Show Detail

    • No:3664
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:20
  • Error Code:[0x89C2]

    «Main Cause» D2 lamp malfunctioned. Lighting circuit malfunctioned. «Countermeasure» Replace the D2 lamp. Click here for Replacement method o... Show Detail

    • No:3663
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:19
  • Error Code:[0x897A]

    «Main Cause» Lamp housing temperature sensor is open. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3662
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8979]

    «Main Cause» Leak sensor is open. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3661
    • Date of Issue:2020/12/02 16:31