FAQ - Frequently Asked Questions
- Products Search > HPLC > Detector
Keyword Search
Frequently Referenced FAQ
FAQ within [ Detector ]
-
«Main Cause» A probable cause is that the D2 lamp heater circuit malfunctioned. «Countermeasure» Click here for Replacement method of Lamp (HPLC... Show Detail
- No:3670
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/07 11:21
-
«Main Cause» A probable cause is that the W lamp malfunctioned. «Countermeasure» Check the total operating time of the lamp, and replace the lamp. If this ... Show Detail
- No:3669
- Date of Issue:2020/12/02 16:31
-
«Main Cause» A probable cause is that the cooling fan blast flow is poor, or the cooling fan or heater circuit malfunctioned. «Countermeasure» Ensure at le... Show Detail
- No:3668
- Date of Issue:2020/12/02 16:31
-
«Main Cause» A probable cause is that the D2 lamp heater circuit malfunctioned. «Countermeasure» Check the total operating time of the lamp, and replace th... Show Detail
- No:3667
- Date of Issue:2020/12/02 16:31
-
«Main Cause» The lamp cover was not in place at instrument startup. «Countermeasure» Mount the lamp cover, and turn the lamp on again. Show Detail
- No:3666
- Date of Issue:2020/12/02 16:31
-
«Main Cause» Probable causes are the W lamp connector is not connected, or the W lamp malfunctioned. «Countermeasure» Check the W lamp connections. Check t... Show Detail
- No:3665
- Date of Issue:2020/12/02 16:31
-
«Main Cause» Probable causes are the D2 lamp connector is not connected, the D2 lamp malfunctioned, or the lighting circuit malfunctioned. «Counter... Show Detail
- No:3664
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/07 11:20
-
«Main Cause» D2 lamp malfunctioned. Lighting circuit malfunctioned. «Countermeasure» Replace the D2 lamp. Click here for Replacement method o... Show Detail
- No:3663
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/07 11:19
-
«Main Cause» Lamp housing temperature sensor is open. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3662
- Date of Issue:2020/12/02 16:31
-
«Main Cause» Leak sensor is open. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3661
- Date of Issue:2020/12/02 16:31
191-200 of Total 266