• Font Size
  • S
  • M
  • L

FAQ within [ Detector ]

191-200 of Total 266

20 / 27page
  • Error Code:[0x58F4] WARNING WL CHECK NO GOOD

    «Main Cause» Wavelength accuracy check failed due to either a cell installation error, a wavelength calibration error, or a sample or bubbles contained in... Show Detail

    • No:4087
    • Date of Issue:2020/12/02 16:45
    • Last Updated:2021/01/20 12:05
  • Error Code:[0x58F7] FATAL MIRROR POSITION

    «Main Cause» Detection of the motor home position failed for the lamp switching mirror control motor. «Countermeasure» Switch the instrument power OFF and c... Show Detail

    • No:4088
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A04]

    «Main Cause» Information in the nonvolatile memory could not be read. «Countermeasure» Calibrate the detector (wavelength calibration, absorbance correction... Show Detail

    • No:4089
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A21]

    «Main Cause» There is a leak in the detector. «Countermeasure» Inspect the detector leak and wipe up any liquid from around the leak sensor. Show Detail

    • No:4090
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A41]

    «Main Cause» Cannot communicate normally between the instrument and workstation. «Countermeasure» Check the communication environment and operating status b... Show Detail

    • No:4091
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A44]

    «Main Cause» Cannot communicate normally between the instrument and workstation. «Countermeasure» Check the communication environment and operating status b... Show Detail

    • No:4092
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A45]

    «Main Cause» Light level exceeded the threshold value due to the long accumulation time. «Countermeasure» Calibrate the accumulation time. Show Detail

    • No:4093
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A51]

    «Main Cause» Communication between the instrument and system controller was interrupted. «Countermeasure» · Check whether PDA detector and system controller... Show Detail

    • No:4094
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A52]

    «Main Cause» Communication between the instrument and workstation was interrupted. «Countermeasure» Check the communication environment and operating status... Show Detail

    • No:4095
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A54]

    «Main Cause» Communication between the instrument and system controller was interrupted. «Countermeasure» · Check whether a LAN cable is disconnected (PDA d... Show Detail

    • No:4096
    • Date of Issue:2020/12/02 16:45

191-200 of Total 266

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.