FAQ - Frequently Asked Questions
- Products Search > HPLC > Detector
Keyword Search
Frequently Referenced FAQ
FAQ within [ Detector ]
-
«Main Cause» Recovery failed when resending transmission data from the instrument to the workstation «Countermeasure» Check the communication environment a... Show Detail
- No:3406
- Date of Issue:2020/12/02 16:09
-
«Main Cause» Light intensity was exceeded because the detector took a long exposure time. «Countermeasure» Calibrate the exposure time. Show Detail
- No:3407
- Date of Issue:2020/12/02 16:09
-
«Main Cause» Communication between the instrument and the workstation was interrupted. «Countermeasure» Check the communication environment and operating st... Show Detail
- No:3408
- Date of Issue:2020/12/02 16:09
-
«Main Cause» The detector cell temperature does not reach the set temperature. «Countermeasure» Replace the temperature control cell. If the error is not ... Show Detail
- No:3409
- Date of Issue:2020/12/02 16:09
-
«Main Cause» The IP address of the instrument is duplicated on a different device. «Countermeasure» Change the IP addresses so there are no duplicates. If ... Show Detail
- No:3410
- Date of Issue:2020/12/02 16:10
-
«Main Cause» There is a fault in the control circuit of the temperature control cell. «Countermeasure» Replace the temperature control cell. Show Detail
- No:3411
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The temperature control cell temperature does not lower to the set temperature even after 5 minutes have elapsed. «Countermeasure» Set the [Cel... Show Detail
- No:3412
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The [Cell Temperature] settings in the method parameters are set to a value outside the setting range. «Countermeasure» Set the [Cell Temperatu... Show Detail
- No:3413
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The leak sensor of the detector is malfunctioning. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales... Show Detail
- No:3414
- Date of Issue:2020/12/02 16:10
-
«Main Cause» The leak sensor of the detector is malfunctioning. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sale... Show Detail
- No:3415
- Date of Issue:2020/12/02 16:10
11-20 of Total 266