  • Error Code:[0x004A] CHECK NO GOOD

    «Main Cause» Wavelength discrepancy exceeds 1nm. «Countermeasure» 1. Cell is not installed properly - Install cell properly 2. Error in wavelength calibrat... Show Detail

    • No:3723
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x0048] ERR EM HOME POS

    «Main Cause» Em grating motor has slipped. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3722
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x0047] ERR EX HOME POS

    «Main Cause» Ex grating motor has slipped. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3721
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x0044] ERR OVER CURRENT

    «Main Cause» Abnormally high current flowing through lamp. «Countermeasure» Check the total lamp illumination time, and replace the lamp. Cli... Show Detail

    • No:3720
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:23
  • Error Code:[0x0043] ERR D2 LAMP

    «Main Cause» Fault in deuterium lamp or in its circuits. «Countermeasure» Check the total lamp illumination time, and replace the lamp. Click... Show Detail

    • No:3719
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:22
  • Error Code:[0x0042] ERR LAMP HOME PO

    «Main Cause» Lamp home position sensor does not operate properly. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3718
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x0041] ERR FIL HOME POS

    «Main Cause» Filter home position sensor does not operate properly. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3717
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x0040] ERR GR HOME POS

    «Main Cause» Motor home position sensor does not operate properly. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3716
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x89DA]

    «Main Cause» Malfunction inside the instrument. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3682
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x89D9]

    «Main Cause» Malfunction inside the instrument. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail