Error Code:[0x00C3] ERR SUP2 HI VOLT
«Main Cause» The actual current passing through the suppressor cartridge did not reach the preset value one minute after the start of regeneration. ... Show Detail
- No:3741
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/04 09:39
-
Error Code:[0x00C2] ERR SUP1 HI VOLT
«Main Cause» The actual current passing through the suppressor cartridge did not reach the preset value one minute after the start of regeneration. ... Show Detail
- No:3740
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/04 09:35
-
Error Code:[0x00C1] ERR VALVE2 HOME
«Main Cause» The home position of the suppressor Ch2 valve failed to detection. «Countermeasure» (1) Disassemble the valve and clean or replace the rotor an... Show Detail
- No:3739
- Date of Issue:2020/12/02 16:31
-
Error Code:[0x00C0] ERR VALVE1 HOME
«Main Cause» The home position of the suppressor Ch1 valve failed to detection. «Countermeasure» (1) Disassemble the valve and clean or replace the rotor an... Show Detail
- No:3738
- Date of Issue:2020/12/02 16:31
-
Error Code:[0x00A4] ERR OVER HEAT (0)
«Main Cause» Electric circuit was damaged. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail
- No:3737
- Date of Issue:2020/12/02 16:31
-
Error Code:[0x00A1] ERR LEAK DETECTED
«Main Cause» Leak detected. «Countermeasure» Inspect and repair plumbing. Wipe away leakage. Show Detail
- No:3736
- Date of Issue:2020/12/02 16:31
-
Error Code:[0x00A0] ERR ZG HOME POS
«Main Cause» The home position of the zero glass drive unit cannot be properly detected. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail
- No:3735
- Date of Issue:2020/12/02 16:31
-
Error Code:[0x0050] DATA NOT EXIST
«Main Cause» Spectrum data was requested for output from a file which dose not contain scanned data. «Countermeasure» Check the file number set when scanni... Show Detail
- No:3726
- Date of Issue:2020/12/02 16:31
-
«Main Cause» Fault in Xe lamp or in its circuits. «Countermeasure» Check the Xe lamp operating time and replace it. Click here for Replacemen... Show Detail
- No:3725
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/07 11:23
-
Error Code:[0x004B] ERR OVER HEAT
«Main Cause» The temperature of the measuring point rises to 63ºC or more. «Countermeasure» Check the operating environment. If you cannot find any problem... Show Detail
- No:3724
- Date of Issue:2020/12/02 16:31