• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Detector ]

161-170 of Total 266

≪ 17 / 27page ≫
  • Error Code:[0x00C3] ERR SUP2 HI VOLT

    «Main Cause» The actual current passing through the suppressor cartridge did not reach the preset value one minute after the start of regeneration. ... Show Detail

    • No:3741
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/04 09:39
  • Error Code:[0x00C2] ERR SUP1 HI VOLT

    «Main Cause» The actual current passing through the suppressor cartridge did not reach the preset value one minute after the start of regeneration. ... Show Detail

    • No:3740
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/04 09:35
  • Error Code:[0x00C1] ERR VALVE2 HOME

    «Main Cause» The home position of the suppressor Ch2 valve failed to detection. «Countermeasure» (1) Disassemble the valve and clean or replace the rotor an... Show Detail

    • No:3739
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x00C0] ERR VALVE1 HOME

    «Main Cause» The home position of the suppressor Ch1 valve failed to detection. «Countermeasure» (1) Disassemble the valve and clean or replace the rotor an... Show Detail

    • No:3738
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x00A4] ERR OVER HEAT (0)

    «Main Cause» Electric circuit was damaged. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3737
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x00A1] ERR LEAK DETECTED

    «Main Cause» Leak detected. «Countermeasure» Inspect and repair plumbing. Wipe away leakage. Show Detail

    • No:3736
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x00A0] ERR ZG HOME POS

    «Main Cause» The home position of the zero glass drive unit cannot be properly detected. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3735
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x0050] DATA NOT EXIST

    «Main Cause» Spectrum data was requested for output from a file which dose not contain scanned data. «Countermeasure» Check the file number set when scanni... Show Detail

    • No:3726
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x004C] ERR LAMP

    «Main Cause» Fault in Xe lamp or in its circuits. «Countermeasure» Check the Xe lamp operating time and replace it. Click here for Replacemen... Show Detail

    • No:3725
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:23
  • Error Code:[0x004B] ERR OVER HEAT

    «Main Cause» The temperature of the measuring point rises to 63ºC or more. «Countermeasure» Check the operating environment. If you cannot find any problem... Show Detail

    • No:3724
    • Date of Issue:2020/12/02 16:31

161-170 of Total 266

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.