Frequently Referenced FAQ

FAQ within [ Detector ]

151-160 of Total 266

≪ 16 / 27page ≫
  • Error Code:[0x871D] ERR CELL TEMP

    «Main Cause» Displayed when the flow cell temperature has risen abnormally. «Countermeasure» If the exhaust vents at the rear and side are obstructed, secur... Show Detail

    • No:3751
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x871C] CHECK NG3

    «Main Cause» No peak was found in the emission wavelength peak detection shown inside the parentheses. «Countermeasure» (1) Fit the flow cell unit accuratel... Show Detail

    • No:3750
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x871B] CHECK NG2

    «Main Cause» No peak was found in the excitation wavelength peak detection shown inside the parentheses. «Countermeasure» (1) Fit the flow cell unit accurat... Show Detail

    • No:3749
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x871A] CHECK NG1

    «Main Cause» Displayed when there is a wavelength discrepancy of grater than 2.0nm in a wavelength check or on wavelength calibration. «Countermeasure» (1) ... Show Detail

    • No:3748
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8719] ERR Hg LAMP

    «Main Cause» Displayed when the Hg lamp cannot be turn on. «Countermeasure» Turn the power OFF and then restart the instrument. If the same error is still d... Show Detail

    • No:3747
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8714] ERR Xe LAMP

    «Main Cause» Displayed when the Xe lamp cannot be turn on. «Countermeasure» Turn the power OFF and back ON, then turn the Xe lamp on. The Xe lamp cannot be ... Show Detail

    • No:3746
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8710] ERR LM HOME

    «Main Cause» Displayed when the home position sensor of the lamp mirror is not operating correctly. «Countermeasure» Turn the power OFF and then restart the... Show Detail

    • No:3745
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x870E] ERR EEPROM WRITE

    «Main Cause» EEPROM write error is occurred. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3744
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x00C5] ERR BG2 TOO HIGH

    «Main Cause» GAIN1/GAIN2 was set to 1 and the background level exceeded 100μScm-1 when autozero was performed. GAIN1/GAIN2 should be 2 or larger if backgroun... Show Detail

    • No:3743
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x00C4] ERR BG1 TOO HIGH

    «Main Cause» GAIN1/GAIN2 was set to 1 and the background level exceeded 100μScm-1 when autozero was performed. GAIN1/GAIN2 should be 2 or larger if backgroun... Show Detail

    • No:3742
    • Date of Issue:2020/12/02 16:31