FAQ within [ Detector ]
-
«Main Cause» The IP address for the PDA detector was changed during use. «Countermeasure» The parent-child relationship might have failed between the system... Show Detail
- No:3810
- Date of Issue:2020/12/02 16:43
-
«Main Cause» The system name setting for the master PDA detector was changed during use. «Countermeasure» The parent-child relationship might have failed be... Show Detail
- No:3811
- Date of Issue:2020/12/02 16:43
-
«Main Cause» The PDA detector system name was changed during use. «Countermeasure» The parent-child relationship might have failed between the system contro... Show Detail
- No:3812
- Date of Issue:2020/12/02 16:43
-
«Main Cause» The PDA detector independently started up the system controller during an analysis. «Countermeasure» Either stop the PDA detector analysis or w... Show Detail
- No:3813
- Date of Issue:2020/12/02 16:43
-
Error Code:[0x58E9] WARNING NOT TEMP-CONT CELL
«Main Cause» The cell temperature could not be controlled to the temperature specified in the method because a cell that does not support temperature-control... Show Detail
- No:4051
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5803] WARNING MICRO BOARD
«Main Cause» The micro cell DET board is not properly recognized. «Countermeasure» Check the cell type. To replace the current DET board with one for a micr... Show Detail
- No:4052
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5821] ERROR LEAK DETECTED
«Main Cause» There is a leak in the detector. «Countermeasure» Inspect the detector leak and wipe up any liquid from around the leak sensor. Show Detail
- No:4053
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5861] FATAL CELL TEMP
«Main Cause» The temperature-controlled cell control circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shi... Show Detail
- No:4054
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5863] WARNING CELL LOW SET TEMP
«Main Cause» The temperature did not decrease to the setting temperature even after five minutes of temperature control. «Countermeasure» Specify a somewhat... Show Detail
- No:4055
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5864] WARNING CELL NOT CONNECTED
«Main Cause» Analysis was started without connecting the flow cell. «Countermeasure» Connect the cell to the main unit. Show Detail
- No:4056
- Date of Issue:2020/12/02 16:45