• Font Size
  • S
  • M
  • L

FAQ within [ Detector ]

151-160 of Total 266

16 / 27page
  • Error Code:[0x80F3]

    «Main Cause» The IP address for the PDA detector was changed during use. «Countermeasure» The parent-child relationship might have failed between the system... Show Detail

    • No:3810
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80F4]

    «Main Cause» The system name setting for the master PDA detector was changed during use. «Countermeasure» The parent-child relationship might have failed be... Show Detail

    • No:3811
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80F5]

    «Main Cause» The PDA detector system name was changed during use. «Countermeasure» The parent-child relationship might have failed between the system contro... Show Detail

    • No:3812
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80F7]

    «Main Cause» The PDA detector independently started up the system controller during an analysis. «Countermeasure» Either stop the PDA detector analysis or w... Show Detail

    • No:3813
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x58E9] WARNING NOT TEMP-CONT CELL

    «Main Cause» The cell temperature could not be controlled to the temperature specified in the method because a cell that does not support temperature-control... Show Detail

    • No:4051
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5803] WARNING MICRO BOARD

    «Main Cause» The micro cell DET board is not properly recognized. «Countermeasure» Check the cell type. To replace the current DET board with one for a micr... Show Detail

    • No:4052
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5821] ERROR LEAK DETECTED

    «Main Cause» There is a leak in the detector. «Countermeasure» Inspect the detector leak and wipe up any liquid from around the leak sensor. Show Detail

    • No:4053
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5861] FATAL CELL TEMP

    «Main Cause» The temperature-controlled cell control circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shi... Show Detail

    • No:4054
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5863] WARNING CELL LOW SET TEMP

    «Main Cause» The temperature did not decrease to the setting temperature even after five minutes of temperature control. «Countermeasure» Specify a somewhat... Show Detail

    • No:4055
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5864] WARNING CELL NOT CONNECTED

    «Main Cause» Analysis was started without connecting the flow cell. «Countermeasure» Connect the cell to the main unit. Show Detail

    • No:4056
    • Date of Issue:2020/12/02 16:45

151-160 of Total 266

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.