Frequently Referenced FAQ

FAQ within [ Detector ]

131-140 of Total 266

≪ 14 / 27page ≫
  • Error Code:[0x2F16]

    «Main Cause» PDA detector: PDA detector failed to connect to the workstation. «Countermeasure» Check the communication environment (LAN cables, hub, etc.). ... Show Detail

    • No:3781
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x2F15]

    «Main Cause» PDA detector: PDA detector failed to connect to the workstation. «Countermeasure» Check the communication environment (LAN cables, hub, etc.). ... Show Detail

    • No:3780
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x2F14]

    «Main Cause» PDA detector: PDA detector failed to connect to the workstation. «Countermeasure» Check the communication environment (LAN cables, hub, etc.). ... Show Detail

    • No:3779
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x2F13]

    «Main Cause» PDA detector: PDA detector failed to connect to the workstation. «Countermeasure» Check the communication environment (LAN cables, hub, etc.). ... Show Detail

    • No:3778
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x2F12]

    «Main Cause» PDA detector: PDA detector failed to connect to the workstation. «Countermeasure» Check the communication environment (LAN cables, hub, etc.). ... Show Detail

    • No:3777
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x2F11]

    «Main Cause» PDA detector: PDA detector failed to connect to the workstation. «Countermeasure» Check the communication environment (LAN cables, hub, etc.). ... Show Detail

    • No:3776
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x87D0] LOW SET TEMP

    «Main Cause» This warning is displayed when the flow cell temperature fails to coincide with temperature settings after undergoing 5 minutes or more of tempe... Show Detail

    • No:3765
    • Date of Issue:2020/12/02 16:32
  • Error Code:[0x874B] ERR OVER HEAT

    «Main Cause» Instrument interior temperature has risen to an abnormal level. «Countermeasure» Check to be sure that the rear fan can move, and that the exh... Show Detail

    • No:3764
    • Date of Issue:2020/12/02 16:32
  • Error Code:[0x8744] ERR OVER CURRENT

    «Main Cause» Abnormally high current flowing through lamp. «Countermeasure» Click here for Replacement method of Lamp (HPLC) Turn the power O... Show Detail

    • No:3763
    • Date of Issue:2020/12/02 16:32
    • Last Updated:2020/12/07 11:24
  • Error Code:[0x8743] ERR D2 LAMP

    «Main Cause» Fault in deuterium lamp or in its circuits. «Countermeasure» Click here for Replacement method of Lamp (HPLC) Turn the power OFF... Show Detail

    • No:3762
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:24