• Font Size
  • S
  • M
  • L

FAQ within [ Detector ]

111-120 of Total 266

12 / 27page
  • Error Code:[0x8714] ERR Xe LAMP

    «Main Cause» Displayed when the Xe lamp cannot be turn on. «Countermeasure» Turn the power OFF and back ON, then turn the Xe lamp on. The Xe lamp cannot be ... Show Detail

    • No:3746
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8719] ERR Hg LAMP

    «Main Cause» Displayed when the Hg lamp cannot be turn on. «Countermeasure» Turn the power OFF and then restart the instrument. If the same error is still d... Show Detail

    • No:3747
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x871A] CHECK NG1

    «Main Cause» Displayed when there is a wavelength discrepancy of grater than 2.0nm in a wavelength check or on wavelength calibration. «Countermeasure» (1) ... Show Detail

    • No:3748
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x871B] CHECK NG2

    «Main Cause» No peak was found in the excitation wavelength peak detection shown inside the parentheses. «Countermeasure» (1) Fit the flow cell unit accurat... Show Detail

    • No:3749
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x871C] CHECK NG3

    «Main Cause» No peak was found in the emission wavelength peak detection shown inside the parentheses. «Countermeasure» (1) Fit the flow cell unit accuratel... Show Detail

    • No:3750
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x871D] ERR CELL TEMP

    «Main Cause» Displayed when the flow cell temperature has risen abnormally. «Countermeasure» If the exhaust vents at the rear and side are obstructed, secur... Show Detail

    • No:3751
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x871E] LAMP COVER OPEN

    «Main Cause» Displayed when lamp cover is not in place. «Countermeasure» Fit lamp cover. When this error occurs, the Xe lamp is forcibly turned off. To reli... Show Detail

    • No:3752
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x871F] ERR PELTIER FAN

    «Main Cause» Displayed when the cooling fan for the Peltier cooling fan has stopped. «Countermeasure» Turn the power OFF and then restart the instrument. Se... Show Detail

    • No:3753
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8720] ERR FAN STOP

    «Main Cause» Displayed when the cooling fan for the Xe lamp's power supply stops. «Countermeasure» Check if the cooling fan is working . If the exhaust vent... Show Detail

    • No:3754
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8721] LOW SET TEMP

    «Main Cause» Displayed if the room temperature exceeds the set temperature for the flow cell by 10ºC or more. «Countermeasure» Set a flow cell temperature t... Show Detail

    • No:3755
    • Date of Issue:2020/12/02 16:31

111-120 of Total 266

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.