• Font Size
  • S
  • M
  • L

FAQ within [ Detector ]

101-110 of Total 266

11 / 27page
  • Error Code:[0x00A1] ERR LEAK DETECTED

    «Main Cause» Leak detected. «Countermeasure» Inspect and repair plumbing. Wipe away leakage. Show Detail

    • No:3736
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x00A4] ERR OVER HEAT (0)

    «Main Cause» Electric circuit was damaged. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3737
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x00C0] ERR VALVE1 HOME

    «Main Cause» The home position of the suppressor Ch1 valve failed to detection. «Countermeasure» (1) Disassemble the valve and clean or replace the rotor an... Show Detail

    • No:3738
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x00C1] ERR VALVE2 HOME

    «Main Cause» The home position of the suppressor Ch2 valve failed to detection. «Countermeasure» (1) Disassemble the valve and clean or replace the rotor an... Show Detail

    • No:3739
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x00C2] ERR SUP1 HI VOLT

    «Main Cause» The actual current passing through the suppressor cartridge did not reach the preset value one minute after the start of regeneration. ... Show Detail

    • No:3740
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/04 09:35
  • Error Code:[0x00C3] ERR SUP2 HI VOLT

    «Main Cause» The actual current passing through the suppressor cartridge did not reach the preset value one minute after the start of regeneration. ... Show Detail

    • No:3741
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/04 09:39
  • Error Code:[0x00C4] ERR BG1 TOO HIGH

    «Main Cause» GAIN1/GAIN2 was set to 1 and the background level exceeded 100μScm-1 when autozero was performed. GAIN1/GAIN2 should be 2 or larger if backgroun... Show Detail

    • No:3742
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x00C5] ERR BG2 TOO HIGH

    «Main Cause» GAIN1/GAIN2 was set to 1 and the background level exceeded 100μScm-1 when autozero was performed. GAIN1/GAIN2 should be 2 or larger if backgroun... Show Detail

    • No:3743
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x870E] ERR EEPROM WRITE

    «Main Cause» EEPROM write error is occurred. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3744
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8710] ERR LM HOME

    «Main Cause» Displayed when the home position sensor of the lamp mirror is not operating correctly. «Countermeasure» Turn the power OFF and then restart the... Show Detail

    • No:3745
    • Date of Issue:2020/12/02 16:31

101-110 of Total 266

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.