Frequently Referenced FAQ

FAQ within [ Detector ]

91-100 of Total 266

≪ 10 / 27page ≫
  • Error Code:[0x58CD] FATAL W LAMP NOT LIT

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Implement corrective measures in the following sequence. (1... Show Detail

    • No:4072
    • Date of Issue:2020/12/02 16:45
    • Last Updated:2020/12/07 11:13
  • Error Code:[0x58CC] ERROR W LAMP NOT CONNECTED

    «Main Cause» The W lamp is not connected or the illumination circuit is not functioning properly. «Countermeasure» Implement corrective measures... Show Detail

    • No:4071
    • Date of Issue:2020/12/02 16:45
    • Last Updated:2020/12/07 11:12
  • Error Code:[0x58CA] ERROR D2 LAMP NOT LIT

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Implement corrective measures in the following se... Show Detail

    • No:4070
    • Date of Issue:2020/12/02 16:45
    • Last Updated:2020/12/07 11:11
  • Error Code:[0x58C9] ERROR W LAMP NOT LIT

    «Main Cause» The W lamp or W lamp illumination circuit is not functioning properly. «Countermeasure» Implement corrective measures in the follow... Show Detail

    • No:4069
    • Date of Issue:2020/12/02 16:45
    • Last Updated:2020/12/07 11:11
  • Error Code:[0x58C7] ERROR D2 LAMP NOT LIT

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail

    • No:4068
    • Date of Issue:2020/12/02 16:45
    • Last Updated:2020/12/07 11:10
  • Error Code:[0x58C6] WORNING LAMP COVER OPEN

    «Main Cause» The lamp housing opened or the cover sensor is not functioning properly. «Countermeasure» Close the lamp housing cover. If that does not resolv... Show Detail

    • No:4067
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x58C4] ERROR W LAMP NOT LIT

    «Main Cause» The W lamp failed to illuminate due to a broken filament or the illumination circuit is not functioning properly. «Countermeasure» ... Show Detail

    • No:4066
    • Date of Issue:2020/12/02 16:45
    • Last Updated:2020/12/07 11:10
  • Error Code:[0x58C3] ERROR D2 LAMP NOT LIT

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail

    • No:4065
    • Date of Issue:2020/12/02 16:45
    • Last Updated:2020/12/07 11:09
  • Error Code:[0x58C2] ERROR D2 LAMP NOT LIT

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail

    • No:4064
    • Date of Issue:2020/12/02 16:45
    • Last Updated:2020/12/07 11:08
  • Error Code:[0x5885] FATAL REAR FAN

    «Main Cause» A rotation problem was detected for the fan on the back of the instrument. «Countermeasure» Switch the instrument power OFF and contact a Shima... Show Detail

    • No:4063
    • Date of Issue:2020/12/02 16:45