• Font Size
  • S
  • M
  • L

FAQ within [ Detector ]

91-100 of Total 266

10 / 27page
  • Error Code:[0x0042] ERR LAMP HOME PO

    «Main Cause» Lamp home position sensor does not operate properly. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3718
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x0043] ERR D2 LAMP

    «Main Cause» Fault in deuterium lamp or in its circuits. «Countermeasure» Check the total lamp illumination time, and replace the lamp. Click... Show Detail

    • No:3719
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:22
  • Error Code:[0x0044] ERR OVER CURRENT

    «Main Cause» Abnormally high current flowing through lamp. «Countermeasure» Check the total lamp illumination time, and replace the lamp. Cli... Show Detail

    • No:3720
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:23
  • Error Code:[0x0047] ERR EX HOME POS

    «Main Cause» Ex grating motor has slipped. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3721
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x0048] ERR EM HOME POS

    «Main Cause» Em grating motor has slipped. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3722
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x004A] CHECK NO GOOD

    «Main Cause» Wavelength discrepancy exceeds 1nm. «Countermeasure» 1. Cell is not installed properly - Install cell properly 2. Error in wavelength calibrat... Show Detail

    • No:3723
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x004B] ERR OVER HEAT

    «Main Cause» The temperature of the measuring point rises to 63ºC or more. «Countermeasure» Check the operating environment. If you cannot find any problem... Show Detail

    • No:3724
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x004C] ERR LAMP

    «Main Cause» Fault in Xe lamp or in its circuits. «Countermeasure» Check the Xe lamp operating time and replace it. Click here for Replacemen... Show Detail

    • No:3725
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:23
  • Error Code:[0x0050] DATA NOT EXIST

    «Main Cause» Spectrum data was requested for output from a file which dose not contain scanned data. «Countermeasure» Check the file number set when scanni... Show Detail

    • No:3726
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x00A0] ERR ZG HOME POS

    «Main Cause» The home position of the zero glass drive unit cannot be properly detected. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3735
    • Date of Issue:2020/12/02 16:31

91-100 of Total 266

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.