FAQ - Frequently Asked Questions
- Products Search > HPLC > Common
Keyword Search
Frequently Referenced FAQ
FAQ within [ Common ]
-
Error Code:[0x5F05] FATAL UNEXPECTED ERROR
«Main Cause» The system was restarted for safety reasons, because a system error occurred. «Countermeasure» Restart the system again. If it occurs frequentl... Show Detail
- No:3855
- Date of Issue:2020/12/02 16:43
-
Is there a strong air flow blowing onto the system from an air conditioner, for example? Use a barrier or other means to block air flow from hitting t... Show Detail
- No:4267
- Date of Issue:2020/12/03 09:24
-
Solvent Delivery Pressure Increases
Inspect the detector flow channels. Inspect/replace the column. Does the pressure increase even after backflushing the outlet tubing? Replace the outl... Show Detail
- No:4273
- Date of Issue:2020/12/03 09:56
- Brached FAQ
-
Tighten the leaking connection. Prevent air flow from directly hitting the instrument. If a detector with temperature-control functionality is used, check ... Show Detail
- No:4274
- Date of Issue:2020/12/03 10:01
- Brached FAQ
-
Poor Area Value Reproducibility
If the solvent delivery pump flow is unstable, then area values and reproducibility are also unstable. Check the solvent delivery pump for pulsation, leakage... Show Detail
- No:4276
- Date of Issue:2020/12/03 10:09
- Brached FAQ
-
Autosamplers include rinsing functionality. Rinse the exterior and interior surfaces of the needle with solvent. Change the type of rinse solution and rins... Show Detail
- No:4278
- Date of Issue:2020/12/03 10:16
- Brached FAQ
-
No power is supplied when power switch is switched ON
Plug in the power cord properly. Replace it with a power cord of the same type. Use a power supply that meets the power supply specifications for the ins... Show Detail
- No:4279
- Date of Issue:2020/12/03 10:18
- Brached FAQ
-
Pump 2-propanol to rinse the check valve. However, if a buffer solution is being used, rinse the check valve with purified water first. Then rinse it by pump... Show Detail
- No:4290
- Date of Issue:2020/12/03 10:43
- Brached FAQ
-
Check the column using known analytical conditions. If the column has deteriorated, replace it. Connect the tubing correctly. Backflush the high pressure... Show Detail
- No:4291
- Date of Issue:2020/12/03 10:46
- Brached FAQ
-
Check the column using known analytical conditions. If the column has deteriorated, replace it. Confirm whether the solvent delivery pump is functioning pr... Show Detail
- No:4292
- Date of Issue:2020/12/03 10:49
- Brached FAQ
11-20 of Total 34