• Font Size
  • S
  • M
  • L

FAQ within [ Column Oven ]

71-76 of Total 76

8 / 8page
  • The retention times for target components are significant...

    Has the column temperature deviated? Check the accuracy of the column oven temperature. If not appropriate, adjust the temperature accuracy. Show Detail

    • No:4266
    • Date of Issue:2020/12/03 09:24
  • [Ready] status never appears

    Close column oven door. Specify an appropriate column temperature-control temperature. The possible temperature-control range varies depending on the room ... Show Detail

    • No:4286
    • Date of Issue:2020/12/03 10:33
    • Brached FAQ
  • Peak retention times vary

    Pump 2-propanol to rinse the check valve. However, if a buffer solution is being used, rinse the check valve with purified water first. Then rinse it by pump... Show Detail

    • No:4290
    • Date of Issue:2020/12/03 10:43
    • Brached FAQ
  • The peak shape is abnormal

    Check the column using known analytical conditions. If the column has deteriorated, replace it. Connect the tubing correctly. Backflush the high pressure... Show Detail

    • No:4291
    • Date of Issue:2020/12/03 10:46
    • Brached FAQ
  • No peaks are output

    Check the column using known analytical conditions. If the column has deteriorated, replace it. Confirm whether the solvent delivery pump is functioning pr... Show Detail

    • No:4292
    • Date of Issue:2020/12/03 10:49
    • Brached FAQ
  • Maintenance for CTO-40C/S

    Consumable Part No. 228-72570-01 Air filter 1. Remove the filter holder (A) on the right side by pressing its right and left knobs (a). 2. ... Show Detail

    • No:4339
    • Date of Issue:2020/12/03 13:45
    • Brached FAQ

71-76 of Total 76

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.