• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Column Oven ]

61-70 of Total 76

≪ 7 / 8page ≫
  • Error Code:[0x562E] ERROR EXTERNAL INPUT

    «Main Cause» Temperature control stopped because an error signal was input via the oven external input terminal. «Countermeasure» Check the connected extern... Show Detail

    • No:3400
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x562D] ERROR GAS DETECTED

    «Main Cause» Temperature control was stopped because the gas sensor inside the oven detected organic solvent leakage. «Countermeasure» (1) If an organic sol... Show Detail

    • No:3399
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x562C] ERROR LEAK DETECTED

    «Main Cause» Temperature control was stopped because leakage was detected in the oven. «Countermeasure» Inspect the leakage from flow channels or other loca... Show Detail

    • No:3398
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x562B] ERROR OVER T.MAX

    «Main Cause» Temperature control was stopped because the oven temperature exceeded the upper limit temperature. «Countermeasure» Set the oven temperature se... Show Detail

    • No:3397
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x562A] FATAL AC INPUT POWER

    «Main Cause» The power supply voltage or frequency was not recognized properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sale... Show Detail

    • No:3396
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5629] ERROR DOOR SENSOR

    «Main Cause» The oven door was left ajar for a long time. «Countermeasure» Make sure the door is securely closed. If it is fully closed, it might be a door ... Show Detail

    • No:3395
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5628] FATAL ROOM TEMP SENSOR

    «Main Cause» A broken or short circuit may have occurred in the room temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shim... Show Detail

    • No:3394
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5627] FATAL AC HEATER

    «Main Cause» The heater is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3393
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5626] FATAL GAS SENSOR

    «Main Cause» A broken circuit may have occurred in the gas sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:3392
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x5625] FATAL OVEN TEMP SENSOR L

    «Main Cause» A broken or short circuit may have occurred in the temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu ... Show Detail

    • No:3391
    • Date of Issue:2020/12/02 15:47

61-70 of Total 76

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.