  • Error Code:[0x9408] FATAL INSIDE FAN

    «Main Cause» The fan inside the compartment may have stopped. «Countermeasure» If tubing or other foreign matter is obstructing fan rotation, switch the ins... Show Detail

    • No:4042
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x9409] FATAL PELTIER UNIT 3

    «Main Cause» The Peltier unit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service represen... Show Detail

    • No:4043
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x940B] ERROR OVER T.MAX

    «Main Cause» Temperature control was stopped because the column oven temperature exceeded the upper limit temperature. «Countermeasure» Set the column oven ... Show Detail

    • No:4044
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x940C] FATAL OVEN TEMP SENSOR H

    «Main Cause» A broken or short circuit may have occurred in the temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu ... Show Detail

    • No:4045
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x940D] FATAL OVEN TEMP SENSOR L

    «Main Cause» A broken or short circuit may have occurred in the temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu ... Show Detail

    • No:4046
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x940E] WARNING LOW SET TEMP

    «Main Cause» The setting is too low, relative to room temperature. «Countermeasure» Check the column oven temperature setting in instrument parameter settin... Show Detail

    • No:4047
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x940F] ERROR WRITE TO CMD

    «Main Cause» Writing to the column management device (CMD) failed. «Countermeasure» Do not disconnect the CMD during an analysis or during writing. If a CMD... Show Detail

    • No:4048
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x9410] FATAL AUX HEATER

    «Main Cause» The auxiliary heater is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service repr... Show Detail

    • No:4049
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x9411] ERROR DOOR SENSOR

    «Main Cause» The column oven door was left ajar for a long time. «Countermeasure» Make sure the door is securely closed. If it is fully closed, it might be ... Show Detail

    • No:4050
    • Date of Issue:2020/12/02 16:45
  • An error message is not displayed even if mobile phase is...

    Is the gas sensor or leak sensor sensitivity inadequate? ·Clean the leak sensor. ·Inspect the gas sensor and leak sensor. If it fails, adjust sensitivi... Show Detail