Frequently Referenced FAQ

FAQ within [ Column Oven ]

51-60 of Total 76

≪ 6 / 8page ≫
  • Error Code:[0x1009]

    «Main Cause» Column oven door was open for too long time while using suppressor. «Countermeasure» Close the door while using suppressor. Show Detail

    • No:3556
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x9413]

    «Main Cause» The set temperature of the column oven is set to 86ºC or higher while the column management device (CMD) is connected. «Countermeasure» Please ... Show Detail

    • No:3548
    • Date of Issue:2020/12/02 16:11
  • Error Code:[0x940A]

    «Main Cause» The room temperature sensor of the instrument has detected an abnormal condition. «Countermeasure» Press the (power switch) to turn the power O... Show Detail

    • No:3495
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9403]

    «Main Cause» The gas sensor of the column oven has detected an abnormal condition. «Countermeasure» Press the (power switch) to turn the power OFF and conta... Show Detail

    • No:3494
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9402]

    «Main Cause» An error was detected with the flow channel selection valve in the column oven. «Countermeasure» Press the (power switch) to turn the power OFF... Show Detail

    • No:3493
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9400]

    «Main Cause» The column oven door is open when the column oven is above 50ºC. «Countermeasure» Clear the error. The alarm sound and error message disappear.... Show Detail

    • No:3492
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9218]

    «Main Cause» The consumable part count data for the autosampler and column oven is corrupted. «Countermeasure» Press the (power switch) to turn the power OF... Show Detail

    • No:3486
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9217]

    «Main Cause» The backup parameters for the autosampler and column oven are corrupted. «Countermeasure» Press the (power switch) to turn the power OFF and co... Show Detail

    • No:3485
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x5630] WARNING LOW SET TEMP

    «Main Cause» The setting is too low, relative to room temperature. «Countermeasure» Check the oven temperature setting in instrument parameter settings. If ... Show Detail

    • No:3402
    • Date of Issue:2020/12/02 15:47
  • Error Code:[0x562F] FATAL LEAK SENSOR

    «Main Cause» The leak sensor cannot be detected properly. A broken or short circuit may have occurred. «Countermeasure» Switch the instrument power OFF and ... Show Detail

    • No:3401
    • Date of Issue:2020/12/02 15:47