FAQ within [ Column Oven ]

51-60 of Total 76

≪ 6 / 8page ≫
  • Error Code:[0x540A] ERROR LEAK DETECTED

    «Main Cause» Temperature control was stopped because leakage was detected in the column oven. «Countermeasure» Inspect the leakage from flow channels or oth... Show Detail

    • No:4032
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x540B] ERROR INT-VALVE2 [0x##]

    «Main Cause» An error was detected in the flow channel selection valve 2 inside the column oven. (The error code "0x##" indicates a valve unit error. Refer t... Show Detail

    • No:4033
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x8633] FATAL GAS SENSOR

    «Main Cause» A broken circuit may have occurred in the gas sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:4034
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x8634] FATAL AC HEATER

    «Main Cause» The heater is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:4035
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x8635] FATAL REAR FAN

    «Main Cause» A fan rotation problem was detected in the power supply heat exhaust fan. «Countermeasure» Switch the instrument power OFF and remove any forei... Show Detail

    • No:4036
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x9401] ERROR GAS DETECTED

    «Main Cause» Temperature control was stopped because the gas sensor inside the column oven detected organic solvent leakage. «Countermeasure» · ... Show Detail

    • No:4037
    • Date of Issue:2020/12/02 16:45
    • Last Updated:2022/05/09 15:32
  • Error Code:[0x9404] FATAL THERMAL FUSE

    «Main Cause» A broken circuit may have occurred in the thermal fuse. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or servic... Show Detail

    • No:4038
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x9405] FATAL PELTIER UNIT 1

    «Main Cause» A failure occurred in the Peltier unit control circuit. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or servic... Show Detail

    • No:4039
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x9406] FATAL PELTIER UNIT 2

    «Main Cause» A broken circuit may have occurred in the Peltier unit. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or servic... Show Detail

    • No:4040
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x9407] FATAL PELTIER FAN

    «Main Cause» A rotation problem was detected for the Peltier unit fan (outside the oven cabinet). «Countermeasure» Switch the instrument power OFF and conta... Show Detail

    • No:4041
    • Date of Issue:2020/12/02 16:45