Error Code:[0x8636] ERR POST COOLER
«Main Cause» The post-column cooler is not rotating correctly. «Countermeasure» If anything is interfering with the fan rotation, turn the power OFF and rem... Show Detail
- No:3628
- Date of Issue:2020/12/02 16:30
-
«Main Cause» Cannot write to oven :Management Device (CMD). «Countermeasure» Once turn off the power, and turn on the power again. If this message still app... Show Detail
- No:3627
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x8630] ERR NO RV-* HOME
«Main Cause» Rotation error in an automatic column-switching valve. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3626
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x0079] COOLER NOT EXIST
«Main Cause» Cooler is not mounted or the cable is snapping. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3581
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x0077] WARNING: LOW SET TEMP
«Main Cause» For CTO-10AC (vp) or CTO-20AC: If the COOLER MODE is "1" or "2", the room temperature is less than -10ºC. When "0" is set, Settings of oven temp... Show Detail
- No:3580
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x0076] ERR LEAK SENS
«Main Cause» Leak sensor malfunctions (probably due to broken/disconnected wire). «Countermeasure» Turn power OFF and contact a Shimadzu sales or service re... Show Detail
- No:3579
- Date of Issue:2020/12/02 16:30
-
«Main Cause» Fan speed does not increase to the set value. «Countermeasure» If the cause is obstruction of the fan by tubing or other object, turn the power... Show Detail
- No:3578
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x0073] ERR LEAK DETECT
«Main Cause» Density of solvent vapor inside column oven has exceeded leak sensor actuation level. «Countermeasure» (1) Mobile phase has leaked inside the o... Show Detail
- No:3577
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x0072] ERR SENSOR L
«Main Cause» Thermosensor readings are abnormally low. The sensor may be disconnected. «Countermeasure» Turn power OFF and contact a Shimadzu sales or servi... Show Detail
- No:3576
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x0071] ERR SENSOR H
«Main Cause» Thermosensor readings are abnormally high. There may be a short circuit in the sensor. «Countermeasure» Turn power OFF and contact a Shimadzu s... Show Detail
- No:3575
- Date of Issue:2020/12/02 16:30