• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Column Oven ]

41-50 of Total 76

5 / 8page
  • Error Code:[0x5400] ERROR INT-VALVE1 [0x##]

    «Main Cause» An error was detected in the flow channel selection valve 1 inside the column oven. (The error code "0x##" indicates a valve unit error. Refer t... Show Detail

    • No:4022
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5401] FATAL ACTIVE PREHEATER

    «Main Cause» A broken circuit may have occurred in the active preheater (optional). «Countermeasure» Switch the instrument power OFF and contact a Shimadzu ... Show Detail

    • No:4023
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5402] FATAL AC INPUT POWER

    «Main Cause» The power supply voltage or frequency was not recognized properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sale... Show Detail

    • No:4024
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5403] ERROR EXTERNAL INPUT

    «Main Cause» Temperature control stopped because an error signal was input via the column oven external input terminal. «Countermeasure» Check the connected... Show Detail

    • No:4025
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5404] ERROR HIGH SET TEMP

    «Main Cause» The temperature setting is too high. The temperature cannot be set higher than 85ºC if a flow channel selection valve or column management devic... Show Detail

    • No:4026
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5405] FATAL LEAK SENSOR

    «Main Cause» The leak sensor cannot be detected properly. A broken or short circuit may have occurred. «Countermeasure» Switch the instrument power OFF and ... Show Detail

    • No:4027
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5406] FATAL ROM FAILURE

    «Main Cause» A system error occurred. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:4028
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5407] FATAL RAM FAILURE

    «Main Cause» A system error occurred. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:4029
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5408] ERROR CMD CHECK SUM

    «Main Cause» Data written in the column management device (CMD) is inconsistent. «Countermeasure» Disconnect or reinitialize the CMD. Show Detail

    • No:4030
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5409] FATAL ROOM TEMP SENSOR

    «Main Cause» A broken or short circuit may have occurred in the room temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shim... Show Detail

    • No:4031
    • Date of Issue:2020/12/02 16:45

41-50 of Total 76

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.