Error Code:[0x5405] FATAL LEAK SENSOR
«Main Cause» The leak sensor cannot be detected properly. A broken or short circuit may have occurred. «Countermeasure» Switch the instrument power OFF and ... Show Detail
- No:4027
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5404] ERROR HIGH SET TEMP
«Main Cause» The temperature setting is too high. The temperature cannot be set higher than 85ºC if a flow channel selection valve or column management devic... Show Detail
- No:4026
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5403] ERROR EXTERNAL INPUT
«Main Cause» Temperature control stopped because an error signal was input via the column oven external input terminal. «Countermeasure» Check the connected... Show Detail
- No:4025
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5402] FATAL AC INPUT POWER
«Main Cause» The power supply voltage or frequency was not recognized properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sale... Show Detail
- No:4024
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5401] FATAL ACTIVE PREHEATER
«Main Cause» A broken circuit may have occurred in the active preheater (optional). «Countermeasure» Switch the instrument power OFF and contact a Shimadzu ... Show Detail
- No:4023
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x5400] ERROR INT-VALVE1 [0x##]
«Main Cause» An error was detected in the flow channel selection valve 1 inside the column oven. (The error code "0x##" indicates a valve unit error. Refer t... Show Detail
- No:4022
- Date of Issue:2020/12/02 16:45
-
Error Code:[0x0070] ERR OVER T.MAX
«Main Cause» Temperature control is turned off because the temperature of the column oven exceeded the maximum temperature. «Countermeasure» Settings the in... Show Detail
- No:3766
- Date of Issue:2020/12/02 16:32
-
Error Code:[0x8650] ERR OVER T.MAX-L
«Main Cause» The column oven temperature of heater L has exceeded the value set for [T.MAX]. «Countermeasure» (1) Lower the column oven temperature set valu... Show Detail
- No:3631
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x8638] ERR DOOR SENS
«Main Cause» The door sensor has failed. «Countermeasure» Turn off the power and contact a Shimadzu sales or service representative. Show Detail
- No:3630
- Date of Issue:2020/12/02 16:30
-
Error Code:[0x8637] ERR COOLBLK FAN
«Main Cause» The block cooling fan is not rotating correctly. «Countermeasure» If anything is interfering with the fan rotation, turn the power OFF and remo... Show Detail
- No:3629
- Date of Issue:2020/12/02 16:30