• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Column Oven ]

31-40 of Total 76

4 / 8page
  • Error Code:[0x5405] FATAL LEAK SENSOR

    «Main Cause» The leak sensor cannot be detected properly. A broken or short circuit may have occurred. «Countermeasure» Switch the instrument power OFF and ... Show Detail

    • No:4027
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5404] ERROR HIGH SET TEMP

    «Main Cause» The temperature setting is too high. The temperature cannot be set higher than 85ºC if a flow channel selection valve or column management devic... Show Detail

    • No:4026
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5403] ERROR EXTERNAL INPUT

    «Main Cause» Temperature control stopped because an error signal was input via the column oven external input terminal. «Countermeasure» Check the connected... Show Detail

    • No:4025
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5402] FATAL AC INPUT POWER

    «Main Cause» The power supply voltage or frequency was not recognized properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sale... Show Detail

    • No:4024
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5401] FATAL ACTIVE PREHEATER

    «Main Cause» A broken circuit may have occurred in the active preheater (optional). «Countermeasure» Switch the instrument power OFF and contact a Shimadzu ... Show Detail

    • No:4023
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5400] ERROR INT-VALVE1 [0x##]

    «Main Cause» An error was detected in the flow channel selection valve 1 inside the column oven. (The error code "0x##" indicates a valve unit error. Refer t... Show Detail

    • No:4022
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x0070] ERR OVER T.MAX

    «Main Cause» Temperature control is turned off because the temperature of the column oven exceeded the maximum temperature. «Countermeasure» Settings the in... Show Detail

    • No:3766
    • Date of Issue:2020/12/02 16:32
  • Error Code:[0x8650] ERR OVER T.MAX-L

    «Main Cause» The column oven temperature of heater L has exceeded the value set for [T.MAX]. «Countermeasure» (1) Lower the column oven temperature set valu... Show Detail

    • No:3631
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8638] ERR DOOR SENS

    «Main Cause» The door sensor has failed. «Countermeasure» Turn off the power and contact a Shimadzu sales or service representative. Show Detail

    • No:3630
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8637] ERR COOLBLK FAN

    «Main Cause» The block cooling fan is not rotating correctly. «Countermeasure» If anything is interfering with the fan rotation, turn the power OFF and remo... Show Detail

    • No:3629
    • Date of Issue:2020/12/02 16:30

31-40 of Total 76

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.